Nonin Onyx II 9550 Скачать руководство пользователя страница 2

Authorized Representative in the European Community

IPX2

Protected against vertically falling water drops when enclosure is tilted up to 15 degrees, per 

IEC 60529.

Installing Batteries

Two 1.5 volt AAA-size batteries power the Onyx

®

 II for about 2,500 spot checks or 21 hours of operation. NONIN® 

recommends using alkaline batteries (included with each new Onyx II). When batteries are low, the numeric displays flash 
once per second. Remove batteries if the device will be stored for more than 30 days. Replace low batteries as soon as 
possible, using the instructions below. 

Note: Rechargeable batteries may be used; however, they require more frequent replacement.

1. Hold the Onyx

®

 II as shown at left, pressing upward and then pulling outward slightly with the thumb to release the 

device’s battery tray.

2. Remove the battery tray and the old batteries, disposing of the batteries properly.
3. Insert two new 1.5 volt AAA-size batteries. Follow the polarity markings (+ and -) as illustrated. 

Proper positioning of the 

batteries is essential for operation.

4. Carefully guide the battery tray back onto the Onyx

®

 II, pressing downward and pushing inward slightly to re-secure the 

battery tray. 

Do not force it into place; it fits only when properly positioned.

5. Insert your finger into the device to verify operation.

Activating the Onyx

®

 II and Verifying Operation

The Onyx

®

 II contains numeric LEDs that display oxygen saturation and pulse rate. A tricolor LED display provides a visual 

indication of the pulse signal quality, while blinking at the corresponding pulse rate. This display changes colors to alert you 
to changes in pulse quality that may affect the readings: 

green

 indicates a good pulse signal, 

yellow

 indicates a marginal pulse 

signal, and 

red

 indicates an inadequate pulse signal.

Activate the Onyx

®

 II by inserting the patient’s finger into the unit. The Onyx

®

 II detects the inserted finger and automatically 

illuminates the displays. Correct positioning of the light emitter and photodetector on the finger is critical for accurate 
measurements. All emitted light must pass through the fingertip.
While on the finger, do not press the Onyx

®

 II against any surface and do not squeeze or hold it together. The internal spring 

provides the correct pressure; additional pressure may cause inaccurate readings.
1. Insert the patient’s finger, nail side up, into the Onyx

®

 II until the fingertip touches the built-in stop guide.

2. Make sure the finger is lying flat (not on its side) and is centered within the device. For best results, keep the Onyx

®

 II at 

the patient’s heart or chest level.

3. If the device does not turn on, remove the finger and wait a few seconds before reinserting it.
When a finger is inserted, the Onyx

®

 II performs a brief startup sequence. Verify that all LEDs illuminate during the startup 

sequence. If any LED is not lit, do not use the Onyx

®

 II; contact NONIN

®

 Customer Support for repair or replacement. After 

the startup sequence, the Onyx

®

 II begins sensing the pulse (indicated by the blinking pulse quality display). Allow the device 

to stabilize, observing about 4 seconds of continuous green-colored pulse quality before relying on the displayed values. It is 
common for the displayed values to fluctuate slightly over a period of several seconds. If the pulse quality display blinks yellow 
or red, try another finger. 
A minus sign (-) appears in the left-most digit of the %SpO

2

 display when the Onyx

®

 II senses that the finger has been 

removed. The last measured SpO

2

 and pulse rate values freeze for 10 seconds, and then “OFF” will appear in the display 

while the device turns off. The device will automatically shut off (to conserve battery life) approximately 10 seconds after the 
finger is removed, or after a 2-minute period of inadequate pulse signals.

Using the Lanyard and Carrying Case

 A lanyard and carrying case are provided for convenience. The Onyx

®

 II will function with or without the lanyard.

If lanyard use is desired, thread the lanyard as shown below.

Onyx

®

 II Care, Maintenance, and Cleaning

The advanced digital circuitry within the Onyx

®

 II requires no calibration or periodic maintenance other than battery 

replacement. Field repair of the Onyx

®

 II circuitry is not possible. Do not attempt to open the Onyx

®

 II case or repair the 

electronics. Opening the case will damage  the Onyx

®

 II and void the warranty. Do not open the Onyx

®

 II more than 90°, and 

do not twist or pull on the device when cleaning.

Cleaning the Inner Surfaces of the Onyx

®

 II

1. Wipe the surfaces with a soft cloth dampened with a mild detergent or isopropyl alcohol solution. If low-level disinfection 

is required, a cloth dampened with 10% bleach / 90% water solution may also be used. Do not use undiluted bleach or 
any cleaning solution other than those recommended here, as permanent damage could result.

2. Dry with a soft cloth, or allow to air dry.

!

Caution

Do not use caustic or abrasive cleaning agents.

3. Ensure that all surfaces are completely dry. 

EC REP

1

&&&

&&&

11

Perfusion Indicator

3

4

Содержание Onyx II 9550

Страница 1: ...ravascular dyes may affect the accuracy of the SpO2 measurement Do not autoclave or immerse this device in liquid or use caustic or abrasive cleaning agents A flexible circuit connects the two halves Do not twist or pull the flexible circuit or overextend the device s spring Do not hang the lanyard from the device s flexible circuit strain relief A functional tester cannot be used to assess the ac...

Страница 2: ...p touches the built in stop guide 2 Make sure the finger is lying flat not on its side and is centered within the device For best results keep the Onyx II at the patient s heart or chest level 3 If the device does not turn on remove the finger and wait a few seconds before reinserting it When a finger is inserted the Onyx II performs a brief startup sequence Verify that all LEDs illuminate during ...

Страница 3: ...o 95 non condensing Altitude Operating Up to 40 000 feet 12 192 meters Hyperbaric Pressure Up to 4 atmospheres Battery Life Continuous Approximately 2 500 spot checks based on 21 hours of operation using two AAA size alkaline batteries calculated at 30 seconds per spot check Battery Life Storage 4 years minimum Classifications per IEC 60601 1 CSA 601 1 UL 60601 1 Degree of Protection Type BF Appli...

Страница 4: ... use in the electromagnetic environment specified below The customer and or user of this device should ensure that it is used in such an environment Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material relative humidity should be at least 30 Electrical Fast Transient Burst ...

Страница 5: ...is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones an...

Отзывы: