Nonin Onyx II 9550 Скачать руководство пользователя страница 1

Instructions for Use

Onyx

®

 II Model 9550 Finger Pulse Oximeter

Indications for Use

The NONIN® Onyx

®

 II Model 9550 Finger Pulse Oximeter is a small, lightweight, portable device indicated for use in 

measuring and displaying functional oxygen saturation of arterial hemoglobin (%SpO

2

) and pulse rate of patients who are well 

or poorly perfused. It is intended for spot-checking of adult and pediatric patients on fingers (other than the thumb) between 
0.3 - 1.0 inch (0.8 – 2.5 cm) thick. 

Caution: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.

Contraindications

Do not use the device in an MRI environment, in an explosive atmosphere, or on infant or neonatal patients.

This device is not defibrillation proof per IEC 60601-1:1990 clause 17h.

Warnings

This device is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of 
assessing clinical signs and symptoms.

The device must be able to measure the pulse properly to obtain an accurate SpO

2

 measurement.  Verify that nothing is 

hindering the pulse measurement before relying on the SpO

2

 measurement. 

Operation of this device below the minimum amplitude of 0.3% modulation may cause inaccurate results. 

General operation of the device may be affected by the use of an electrosurgical unit (ESU).

The use of accessories other than those specified in these instructions may result in increased electromagnetic emission 
and/or decreased immunity of this device. 

This device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the 
device should be observed carefully to verify normal operation.

Cautions

This device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Significant 
levels of dysfunctional hemoglobin may affect measurement accuracy.

This device has no audible alarms and is intended only for spot-checking. 

Inspect the sensor application site at least every 6 to 8 hours to ensure correct sensor alignment and skin integrity. Patient 
sensitivity to sensors may vary due to medical status or skin condition.

Fluctuating or very bright light, moisture, blood pressure cuffs, infusion lines, venous pulsations, insufficient pulse signals, 
anemia, arterial catheters, nail polish, and/or artificial nails may degrade the device’s performance.

The device may not work when circulation is reduced. Warm or rub the finger, or re-position the device.

This device’s display will go blank after 30 seconds of no readings or poor readings.

In some circumstances, the device may interpret motion as good pulse quality. Minimize patient motion as much as 
possible.

Cardiogreen and other intravascular dyes may affect the accuracy of the SpO

2

 measurement.

Do not autoclave or immerse this device in liquid or use caustic or abrasive cleaning agents.

A flexible circuit connects the two halves. Do not twist or pull the flexible circuit or overextend the device’s spring.

Do not hang the lanyard from the device’s flexible circuit/strain relief.

A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor or sensor.

This equipment complies with IEC 60601-1-2:2001 for electromagnetic compatibility for medical electrical equipment and/
or systems. This standard is designed to provide reasonable protection against harmful interference in a typical medical 
installation. However, because of the proliferation of radio-frequency transmitting equipment and other sources of 
electrical noise in healthcare and other environments, it is possible that high levels of such interference due to close 
proximity or strength of a source might disrupt the performance of this device. Medical electrical equipment needs special 
precautions regarding EMC, and all equipment must be installed and put into service according to the EMC information 
specified in this manual. 

Portable and mobile RF communications equipment can affect medical electrical equipment.

Batteries may leak or explode if used or disposed of improperly. Remove batteries if the device will be stored for more 
than 30 days. Do not use different types of batteries at the same time. Do not mix fully charged and partially charged 
batteries at the same time. These actions may cause the batteries to leak.

Follow local, state and national governing ordinances and recycling instructions regarding disposal or recycling of the 
device and device components, including batteries.

In compliance with the European Directive on Waste Electrical and Electronic Equipment (WEEE) 2002/96/EC, do not 

dispose of this product as unsorted municipal waste. This device contains WEEE materials; please contact your distributor 

regarding take-back or recycling of the device. If you are unsure how to reach your distributor, please call NONIN® for 

your distributor’s contact information.

Symbols

Symbol

Definition of Symbol

Consult Instructions for Use

  

Caution! 

CE Marking indicating conformance to EC Directive No. 93/42/EEC concerning medical devices

Type BF Applied Part (patient isolation from electrical shock)

Not for Continuous Monitoring (no alarm for SpO

2

)

SN

Serial Number

Battery Orientation

UL Mark for Canada and the United States with respect to electric shock, fire, and mechanical 

hazards only in accordance with UL 60601-1 and CAN/CSA C22.2 No. 601.1.
Lot Number
Non-ionizing electromagnetic radiation. Equipment includes RF transmitters. Interference may 

occur in the vicinity of equipment marked with this symbol.

Indicates separate collection for electrical and electronic equipment (WEEE).

0123

'&

%&

11

!

0123

CL

ASSIFIED

US

C

UL

Содержание Onyx II 9550

Страница 1: ...ravascular dyes may affect the accuracy of the SpO2 measurement Do not autoclave or immerse this device in liquid or use caustic or abrasive cleaning agents A flexible circuit connects the two halves Do not twist or pull the flexible circuit or overextend the device s spring Do not hang the lanyard from the device s flexible circuit strain relief A functional tester cannot be used to assess the ac...

Страница 2: ...p touches the built in stop guide 2 Make sure the finger is lying flat not on its side and is centered within the device For best results keep the Onyx II at the patient s heart or chest level 3 If the device does not turn on remove the finger and wait a few seconds before reinserting it When a finger is inserted the Onyx II performs a brief startup sequence Verify that all LEDs illuminate during ...

Страница 3: ...o 95 non condensing Altitude Operating Up to 40 000 feet 12 192 meters Hyperbaric Pressure Up to 4 atmospheres Battery Life Continuous Approximately 2 500 spot checks based on 21 hours of operation using two AAA size alkaline batteries calculated at 30 seconds per spot check Battery Life Storage 4 years minimum Classifications per IEC 60601 1 CSA 601 1 UL 60601 1 Degree of Protection Type BF Appli...

Страница 4: ... use in the electromagnetic environment specified below The customer and or user of this device should ensure that it is used in such an environment Electrostatic Discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material relative humidity should be at least 30 Electrical Fast Transient Burst ...

Страница 5: ...is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones an...

Отзывы: