![Nonin Avant 4000 Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/nonin/avant-4000/avant-4000_operators-manual_1700077036.webp)
33
Dimensions
5.5” H x 7.25” W x 4.5” D
Weight
2.2 lbs with battery
Memory
33.5 hours minimum
AVANT 4000 DIGITAL PULSE OXIMETRY SYSTEM
Temperature
Operating
32° to +122°F (0° to +50°C)
Storage/Transportation
-22° to +122°F (-30° to +50°C)
Operating Altitude
Up to 40,000 feet
Hyperbaric Pressure
Up to 4 atmospheres
Humidity
Operating
10% to 90% relative humidity, noncondensing
Storage/Transportation
10% to 95% relative humidity, noncondensing
*If transferred from a non-operating temperature and/or
humidity condition, allow at least one hour of
stabilization time before use.
Antenna Type
Inverted F type antenna
Antenna Gain
+2 dB (typ.), +3 dB (max.)
TRANSMITTER
Bluetooth Compliance:
Version 1.1
Operating Frequency:
2.4 to 2.4835 GHz
Output Power:
< 1.1 mW
Operating Range:
10-meter radius indoors
Network Topology:
Point-to-Point
Operation:
Slave: Model 4100; Master: Model 4000
Antenna Type:
Internal
Classifications per IEC 60601- 1 / CSA601.1 / UL2601-1
Type of Protection
Class I (when on AC power with 300PS-UNIV battery
charger)
Internally powered (on battery power)
Degree of Protection
Type BF-Applied Part
Mode of Operation
Continuous
a.
S.D. (Standard Deviation) is a statistical measure; up to 32% of the readings may fall outside
these limits.