Alarm Summary
If patient SpO
2
or pulse readings are equal to or above the upper alarm limit, or if
they are equal to or below the lower alarm limit, the display unit will signal an alarm.
Silencing Alarms
Momentary (2-Minute) Alarm Silence
To silence alarms for two minutes, press the Alarm Silence button.
Continuous Alarm Silence
In order to permanently silence all alarms, DIP switch 2 must be placed in the UP
position. This allows the alarm volume to be set to zero.
The Alarm Silence LED will
remain illuminated when the alarm volume is set to zero.
Refer to “DIP Switches” for more
information.
High Priority Alarm Description
Factory
Default
Adjustment Range
Step Value
SpO
2
Upper Alarm Limit
Off
Off, 80 to 100
1% SpO
2
SpO
2
Lower Alarm Limit
80%
Off, 50 to 95
1% SpO
2
Pulse Upper Alarm Limit
200 BPM
Off, 75 to 275
5 BPM
Pulse Lower Alarm Limit
50 BPM
Off, 30 to 110
5 BPM