26

Alarms and Limits

Adjusting Volume and Alarm Limits 

Follow the instructions below to review, set, or adjust volumes and alarm limits while 
operating the Avant 4000 System. 

Do not use these instructions to store adjusted 

limits. 

For more information about permanently storing adjusted settings, refer to 

“Default Settings.”

IMPORTANT! Alarm limits reset themselves to default values each time the 

unit is powered up.

Reviewing, Setting, or Adjusting SpO

2

 and/or Pulse Alarm Limits

1.

Ensure that the display unit is on.

2.

Press the Alarm Limits button.

• Notice that the upper round LED is illuminated to the left of the Alarm 

Limits button. This indicates the upper alarm limit, while the lower LED 
indicates the lower alarm limit.

• Notice that the current setting appears in the %SpO

2

 display. 

• Continue to press the Alarm Limits button until the alarm limit you want to 

adjust is displayed. 

• The Alarm Limits button can also be cycled to exit Set/Change mode, or 

Set/Change mode will exit automatically after ten seconds with no activity.

3.

Ensure that the appropriate upper or lower Alarm Limit LED is illuminated, 
and that the alarm limit you want to change is displayed.

4.

Press the Plus (+) or Minus (-) buttons to adjust the values as desired. Once 
adjusted, Set/Change mode will exit automatically after ten seconds with no 
activity, and the new limits will be active.

Reviewing, Setting, or Adjusting Pulse and/or Alarm Volumes

1.

Ensure that the display unit is on.

2.

Press the Volume button once to change the alarm volume, or twice to change 
the pulse volume.

• After pressing the Volume button once, notice that the Alarm Volume LED 

appears, and the current setting appears in the Pulse Rate display area. 

• After pressing the Volume button twice, notice that the Pulse Volume LED 

appears, and the current setting appears in the Pulse Rate display area.

• The Volume button can also be cycled to exit Set/Change mode, or Set/

Change mode will exit automatically after ten seconds with no activity.

3.

Use the Plus (+) or Minus (-) buttons to adjust the alarm or pulse volumes as 
desired. Once adjusted, Set/Change mode will exit automatically after ten 
seconds with no activity, and the new limits will be active.

Содержание Avant 4000

Страница 1: ...Operator s Manual Avant 4000 Digital Pulse Oximetry System with Bluetooth Wireless Technology Avant 4000 0123 English...

Страница 2: ...ithout notice or obligation Nonin Medical Inc 2605 Fernbrook Lane North Plymouth Minnesota 55447 4755 USA 1 763 553 9968 800 356 8874 USA and Canada Fax 1 763 553 7807 mail nonin com www nonin com Aut...

Страница 3: ...14 Verifying System Operation 16 Device Pairing 17 Default Settings 18 Accessing User Functions 19 Display Unit DIP switches 22 Care and Maintenance 23 Alarms and Limits 24 Alarms and Informational To...

Страница 4: ...materials Type BF Applied Part Patient isolation from electrical shock UL Mark for Canada and the United States with respect to electric shock fire and mechanical hazards only in accordance with UL 2...

Страница 5: ...angulation All parts and accessories connected to the serial port of this system must be certified according to IEC Standard EN 60950 or UL 1950 for data processing equipment To avoid the risk of conf...

Страница 6: ...te at least every 6 to 8 hours to ensure correct sensor alignment and skin integrity Patient sensitivity to sensors may vary due to medical status or skin condition Do not place liquids on top of this...

Страница 7: ...g measures 1 Reorient or relocate the receiving antenna 2 Increase the distance between the equipment and the receiver 3 Connect the equipment to an outlet on a circuit different from the outlet where...

Страница 8: ...es 8000AA WO finger clip sensor Operator s manual Universal Desktop Battery Charger with IEC320 Connector 3 wrist bands Indications for Use The NONIN Avant 4000 Digital Pulse Oximetry System is indica...

Страница 9: ...ate and plethysmographic data to be transmitted through a Bluetooth radio to a compatible Bluetooth enabled device Nonin s system removes the connection from the sensor cable to the display unit givin...

Страница 10: ...necessary to operate the device without batteries audible alarms and memory functions may not be available DO NOT use the device without batteries when patient safety relies upon audible alarms NOTE...

Страница 11: ...8 Using the Avant 4000 Digital Pulse Oximetry System Installing Batteries in the Patient Module...

Страница 12: ...ED 7 Numeric LEDs for Pulse Rate data 8 Alarm Silence Button 9 Alarm Limits Button and Indicator 10 Plus Button 11 Minus Button 12 Time LED 13 Month LED 14 Day LED 15 Year LED 16 Pulse Volume LED 17 A...

Страница 13: ...s these LEDs display in green For high priority patient alarms the corresponding values are displayed in red blinking fast The values are displayed in amber for medium priority alarms and when reviewi...

Страница 14: ...lect from multiple wrist worn patient modules during pairing Memory Button This button is used to enter Memory Playback mode Bluetooth Control Button Press and hold this button while turning on the di...

Страница 15: ...by the oximeter The height of the Pulse Strength Bargraph LED is proportional to the pulse signal and the color is determined by pulse strength Green a good pulse strength Amber a marginal pulse stren...

Страница 16: ...l it remains lit steadily AC Power Adapter LED This green LED is displayed when an external power supply is providing power to the display unit Battery LED This amber LED indicates a marginal battery...

Страница 17: ...in the display unit and wrist worn patient module WARNING The battery pack must be installed at all times while the device is operating even when operating on AC power If it is necessary to operate th...

Страница 18: ...n of the display unit and pair the display unit with the desired wrist worn patient module See Verifying System Operation and Device Pairing for more information 6 Secure the wristband to the patient...

Страница 19: ...he next page The Main Alarm LED and Pulse Oximeter LED blink amber until the oximeter produces valid readings Use the procedure below to monitor SpO2 and pulse rate readings in order to verify that th...

Страница 20: ...mum of five pairable patient modules will be displayed separately with the device identification number appearing in the SpO2 and Pulse Rate display areas A unique device identification number is incl...

Страница 21: ...olume adjustments made while the device is operating in Factory Default Setting will be lost when the device is shut down The last adjusted values can be recalled by pressing and holding both the Alar...

Страница 22: ...Check the battery charge Press the ON STANDBY button while the unit is on Battery charge is displayed in 10 increments for 4 seconds in green on the Pulse Strength Bargraph Mute the audible alarms 2...

Страница 23: ...s Press the Alarm Limits button to step through the Limits menu Use the Plus or Minus buttons to adjust alarm limits Set pulse and alarm volumes Press the Volume button to select pulse or alarm volume...

Страница 24: ...with the device identification number appearing in the pulse rate and SpO2 display areas When the desired patient module s identification number is displayed pressing the Bluetooth Control button com...

Страница 25: ...me cannot be disabled Switch 3 Date Format Up International Date format Day Month Year Down U S A Date format Month Day Year Switch 4 Data Output Up Real time once per second data output Down Print on...

Страница 26: ...bleshooting Cleaning the Avant 4000 Digital Pulse Oximetry System Clean all system components with a soft cloth dampened with isopropyl alcohol Do not pour or spray any liquids onto components and do...

Страница 27: ...alarms are sounded as follows beep beep beep short pause beep beep 10 second pause Medium Priority Alarms Medium priority alarms are those that signal potential problems with the equipment or other no...

Страница 28: ...rder to permanently silence all alarms DIP switch 2 must be placed in the UP position This allows the alarm volume to be set to zero The Alarm Silence LED will remain illuminated when the alarm volume...

Страница 29: ...lly after ten seconds with no activity 3 Ensure that the appropriate upper or lower Alarm Limit LED is illuminated and that the alarm limit you want to change is displayed 4 Press the Plus or Minus bu...

Страница 30: ...steps 1 Turn the unit off and then back on again to remove the error code 2 If the error persists disconnect all power AC and battery and then reconnect the power and turn the unit back on 3 If the er...

Страница 31: ...is stored in memory starting a new recording session See Displays Indicators and Controls for more information about setting the clock Only recording sessions greater than one minute in length are st...

Страница 32: ...se Rate LED display areas This message signals that the device is in Playback mode 3 The PLy bAC message will disappear when memory playback is complete Memory playback may take up to 8 minutes depend...

Страница 33: ...em in one of the following formats If DIP switch 8 is in the DOWN position the data will be displayed as follows SPO2 XXX HR YYY where XXX is the SpO2 value and YYY is the heart rate value If DIP swit...

Страница 34: ...real time once per second data output is available via the RS232 connector port using a null modem cable When DIP switch 4 is in the DOWN position default the Print on Demand mode is activated allowi...

Страница 35: ...anometers 3 mw nominal Internal Power Battery Two 1 5 volt AA batteries Operating Life minimum 120 hours of continuous operation with new batteries Storage Life 10 months Weight 4 4 ounces with batter...

Страница 36: ...w at least one hour of stabilization time before use Antenna Type Inverted F type antenna Antenna Gain 2 dB typ 3 dB max TRANSMITTER Bluetooth Compliance Version 1 1 Operating Frequency 2 4 to 2 4835...

Страница 37: ...nector Contact your distributor or NONIN for options Cord Set Charger External Cables UNI RS232 RS 232 Cable Pulse Oximeter Sensors 8000AA WO Adult Articulated Finger Clip Sensor 8000J WO Adult Flex P...

Страница 38: ...sive remedy by the purchaser hereunder for any Avant 4000 Digital Pulse Oximetry System delivered to the purchaser which is found to be defective in any manner whether such remedies be in contract tor...

Страница 39: ...Reposition the digit or insert a different digit and keep the sensor motionless for at least 10 seconds Warm the digit by rubbing or covering with a blanket Position the sensor at a different site NOT...

Страница 40: ...to the 4100 Check the 4100 s batteries Check the sensor for any visible signs of deterioration Contact your distributor or NONIN Customer Support Frequent or steady pulse quality indication There is e...

Страница 41: ...connections and retry Verify that the 4100 is paired with the display Verify that the Connection Status LED is green on both the display unit and the patient module Contact your distributor or NONIN...

Страница 42: ...IP switch 2 to the DOWN position if you desire audible alarms The system is not functioning correctly Contact your distributor or NONIN Customer Support The devices will not pair The sensor has not be...

Отзывы: