Koheras HARMONIK
Fiber Coupler Installation Guide

 5  

Step 6: Adjust the focus

a

While monitoring the power meter, rotate the eccentric key until 
the maximum signal is attained.

b

Iteratively repeat steps “5g” (tilt optimization) and “6a” (focus 
optimization) until the maximum signal cannot be increased 
further.

c

Tighten the two grub screws holding the collimator lens.

Step 7: Adjust the fine tilt

a

Once more, check if you can enhance coupling efficiency by using 
the three tilt adjustment screws.

b

Finally, lock the tilt adjustment by tightening the three tilt locking 
screws. Besides their locking function, the tilt locking screws can 
also adjust the fine tilt. 

c

While tightening the first two tilt locking screws, the power meter 
signal typically drops significantly. However, when tightening the 
third screw, the maximum signal should recover.

Step 8: Final check

a

While observing the fiber output power, gently press against the 
cladding stripper of the fiber from the side.

b

When pressing against the cladding stripper in different directions, 
the signal normally drops.

c

Release the cladding, the signal should recover to its maximum 
value.

d

In case the signal increases while pressing against the connector 
in a specific direction, the adjustment is not at its optimum. The 
corresponding locking screw has to be fine tuned further, until the 
fiber output power decreases symmetrically in each push 
direction.

c

“grub screws”

“tilt locking screws”

“cladding”