Nivetec Signet 2760 DryLoc Скачать руководство пользователя страница 5

5

Signet 2760 DryLoc pH/ORP Preampli

fi

 er

Use standard installation hardware to 
connect the submersible 2760 directly 
to external equipment.

½  in. NPT threads

The 8052-1 NPT Mount Junction 
box connects to 

3

/

4

 in. pipe or 

conduit and provides convenient 
wiring termination.

The 8050-1 Universal Mount 
junction box mounts 

fl

 at onto a 

wall or can be strapped to a post 
or pipe.

1.

  Insert any DryLoc electrode into base of the 2760. Lubricate O-rings with a non-petroleum based, viscous lubricant (grease) 

compatible with the system.

2.

  Thread retaining cap over electrode and hand-tighten onto 2760.

 

•   Keep the DryLoc electrical interconnection between electrode and preampli

fi

 er dry and clean at all times.

3.

 Attach 

3

/

4

 in. watertight pipe to the top of the 2760. Secure the threaded connection to prevent any leakage.

 

•   Fill the lower 75-100 mm (3 to 4 in.) of conduit or extension pipe with a 

fl

 exible sealant such as silicone to prevent liquid 

accumulation on the back seal of the 2760.

OR

OR

The submersible 2760 can be installed in a tank or vessel if conditions are within pressure and temperature speci

fi

 cations.

The 2760 is able to withstand 

fl

 uid temperatures up to 85 °C (185° F), but the electrode tolerances vary. Check the speci

fi

 cations of 

each component before installation.

Fill with

75-100 mm

(3 to 4 in.)

of sealant

Fill with

75-100 mm

(3 to 4 in.)

of sealant

Fill with

75-100 mm

(3 to 4 in.)

of sealant

Lubricate O-ring

1

2

55 mm
(2.16 in.)

55 mm
(2.16 in.)

Fill with
75-100 mm
(3 to 4 in.)
of sealant

3

Submersible Sensor Assembly and Installation

Содержание Signet 2760 DryLoc

Страница 1: ...H and ORP electrodes to supply an amplified mV analog signal to Signet instruments 3 2760 090 3 2760 090 Rev 9 01 18 Description Table of Contents English Deutsch Français Español Product Manual Warranty Statement 2 Product Registration 2 Safety Information 2 Chemical Compatibility 2 Specifications 2 Dimensions 3 In Line Assembly and Installation 3 4 Wet Tap Assembly and Installation 4 Submersible...

Страница 2: ...on warranty repairs being returned must include a fully completed Service Form and goods must be returned to your local GF Sales office or distributor Product returned without a Service Form may not be warranty replaced or repaired Signet products with limited shelf life e g pH ORP chlorine electrodes calibration solutions e g pH buffers turbidity standards or other solutions are warranted out of ...

Страница 3: ...ocked When fully assembled the 2760 will stand about 100 mm 4 in above the installation fitting Lock Unlock 100 mm 4 in Lubricate O rings 1 2 3 4 5 2760 In Line models used with 2764 2767 or 2774 2777 series electrodes BEFORE YOU START 2764 2767 Install a 1 in NPT pipe tee into the process line 2774 2777 Install a in reducing tee into the process line 1 Wrap the process threads on the electrode wi...

Страница 4: ... tee in the process line DO NOT install in a inch tee or probe will be damaged 1 Wrap the process threads on the electrode with an appropriate sealant example PTFE tape to create a water tight connection 2 Lubricate O rings with a non petroleum based viscous lubricant grease compatible with the system 3 Remove the threaded nut from the preamplifier and place it around the electrode from the bottom...

Страница 5: ...ier dry and clean at all times 3 Attach 3 4 in watertight pipe to the top of the 2760 Secure the threaded connection to prevent any leakage Fill the lower 75 100 mm 3 to 4 in of conduit or extension pipe with a flexible sealant such as silicone to prevent liquid accumulation on the back seal of the 2760 OR OR The submersible 2760 can be installed in a tank or vessel if conditions are within pressu...

Страница 6: ...mV Input Brown wire Brown mV 4 20 mA Red V Black V White T Green T Connect this terminal to Earth GND to reduce electrical interference White Temp Green Temp Blue Iso Gnd Brown mV input Shield Black V Red V 12 11 10 9 8 7 6 5 18 17 16 15 14 13 12 11 16 15 14 13 12 11 10 9 3 8750 1 3 8750 2 3 8750 3 ProcessPro pH ORP Transmitter 3 2760 11 3 2760 21 Any of these electrodes can be connected to the Si...

Страница 7: ...le 3 2760 1 3 2760 2 DryLoc Preamplifier In Line 3 2760 11 3 2760 21 White Temperature Green Temperature Blue mV Reference Shield Earth Ground Black V Red V Brown Active mV pH ORP Signet Combination pH Electrodes 3 2724 10 3 KΩ NPT 3 2724 11 3 KΩ ISO 3 2726 10 3 KΩ NPT 3 2726 11 3 KΩ ISO 3 2726 HF 10 HF Resist 3 KΩ NPT 3 2726 HF 11 HF Resist 3 KΩ ISO 3 2726 LC 10 Low Cond 3 KΩ NPT 3 2726 LC 11 Low...

Страница 8: ...ings 3 2726 10 159 001 555 Bulb glass pH Electrode 3 KΩ RTD in NPT or Signet fittings 3 2726 11 159 001 556 Bulb glass pH Electrode 3 KΩ RTD ISO 7 1 R or Signet fittings 3 2726 HF 10 159 001 551 Bulb glass pH Electrode 3 KΩ RTD HF resistant in NPT or Signet fittings 3 2726 HF 11 159 001 552 Bulb glass pH Electrode 3 KΩ RTD HF resistant ISO 7 1 R or Signet fittings 3 2726 LC 10 159 001 559 Bulb gla...

Отзывы: