4
Signet 2760 DryLoc pH/ORP Preampli
fi
er
The 2760 DryLoc Preampli
fi
er can be used to connect Signet 275X Wet-Tap electrodes to a pH or ORP instrument.
1. Select the process
fi
tting.
• Select a low pro
fi
le installation saddle (customer-supplied) with a 1½ in. or 2 in. outlet (depending on 3719 model ordered).
• The 3719 is available with NPT or ISO thread options.
• The 3719 can also be adapted to
fi
t into tee
fi
ttings. Refer to the 3719 pH Wet-Tap instructions for detailed information.
2. Install the Wet-Tap and the electrode.
• Install the 3719 into the
fi
tting, then insert the electrode into the assembly.
3. Attach the preampli
fi
er to the electrode.
• Attach the 3-2760-11 or -21 Preampli
fi
er to the DryLoc
fi
tting at the top of the electrode.
4. Connect the 2760 Preampli
fi
er to the instrument.
• The 2760 Preampli
fi
er signal can be routed to any 5700 or 8750 pH/ORP instrument (see Pg. 5).
2760 Submersible models used as in-line system:
The Submersible 2760 design can also be used in-line when
using the 2724-2764, 2764-2767 series or the 2774-2777
series electrodes.
BEFORE YOU START:
2764-2767: Install a 1 in. NPT pipe tee in the process line.
2774-2777 and 2724-2726: Install a ¾-in. reducing tee in
the process line.
DO NOT
install in a ¾-inch tee or probe
will be damaged.
1. Wrap the process threads on the electrode with an
appropriate sealant (example: PTFE tape) to create a
water-tight connection.
2. Lubricate O-rings with a non-petroleum based, viscous lubricant (grease) compatible with the system.
3. Remove the threaded nut from the preampli
fi
er and place it around the electrode from the bottom as illustrated.
4.
HAND-TIGHTEN
the electrode into the installation
fi
tting.
DO NOT USE TOOLS!
5. Seat the 2760 DryLoc connector onto the top of the electrode.
6. Lift the threaded nut up and thread it hand-tight onto the preampli
fi
er.
When the 2760 Preampli
fi
er is fully assembled with one of the threaded electrodes (276X and 277X models) it will stand
approximately 178 mm (7 in.) from the process threads to the top of the preampli
fi
er.
178 mm
(7 in.)
2760 Preamplifier
Electrode
Threaded nut
Lubricate O-ring
2
1
3
4
5
6
• The 2760 Preampli
fi
er signal can be routed to a Signet 5700 ProPoint pH Monitor, or to any model of the 3-8750 pH Transmitter.
• If the preampli
fi
er cable must be extended beyond the standard 4.6m (15 ft) that is supplied with the 2760, use a junction box to
terminate the wires.
Note:
Do not extend the 2760 cable more than 120 m (400 ft).
In-line Assembly and Installation
Wet-Tap Assembly and Installation
RELA 1
2
4
Lubricate O-rings
Lubricate O-rings
1.
2.
3.
4.