TRANSIMITTER
27
CONVERT PARTS
TM
Http: // www.flowmeter.cc
CONVERT
Freq Max(Hz)
2000
<-ENT ESC->
3000
4
3
Click “ ” and “ ”key to set
“Freq Max(Hz)”,Click “ ” show
confirm menu
1
2
Steps
Operation instructions
Interface show
Click “ ” into “Freq Max(Hz)”
Click “ ” key to confirm and return
configuration menu(click” ” give
up modify
Click “ ” three times from configuration menu to measurement
menu, also you can continue other operation.
Freq Max(Hz)
2000
Max: 5000.0
Min: 100.0
2000
Freq Max(Hz)
2000
Max: 5000.0
Min: 100.0
3000
9.) Liter/pulse(L/P)
Current Flow(L/s)
Output Frq (Hz) =
Liter/pulse(L/P)
Liter/pulse(L/P)
When the Liter/ pulse = 0.0, the case "frequency cap Hz" setting determines the
frequency of the output
When the Liter/pulse >0.0, the setting of L/P determines the frequency output
Scale corresponding to the current flow of output frequency
Current Flow(m3/h) /3.6
Steps:
measurement mode
→
system
→
Pulse Output
→
Liter/pulse(L/P)
1
Steps
Operation instructions
Interface show
Click “ ” into “
Liter/pulse
”
Liter/pulse
0.0000
Max: -------
Min: 0.00555
0.0000