24
—
Test frequency
(MHz)
Band
(MHz)
Immunity test levels
Home healthcare environment
385
380-390
Pulse modulation 18Hz, 27V/m
450
430-470
FM, ±5kHz deviation, 1kHz sine, 28V/m
710
704-787
Pulse modulation 217Hz, 9V/m
745
780
810
800-960
Pulse modulation 18Hz, 28V/m
870
930
1720
1700-1990
Pulse modulation 217Hz, 28V/m
1845
1970
2450
2400-2570
Pulse modulation 217Hz, 28V/m
5240
5100-5800
Pulse modulation 217Hz, 9V/m
5500
5785
Table 4 – Input a.c. power Port
Phenomenon
Basic EMC
standard
Immunity test levels
Home healthcare environment
Electrical fast
transients/burst
IEC 61000-4-4
±2 kV
100kHz repetition frequency
Surges
Line-to-line
IEC 61000-4-5
±0.5 kV, ±1 kV
Conducted
disturbances induced
by RF fields
IEC 61000-4-6
3V, 0.15MHz-80MHz
6V in ISM bands and amateur radio bands between
0.15MHz and 80MHz
80%AM at 1kHz