Maintenance
48
XTM0011-F2
XT V 160 X-ray System
Emergency
Switch OFF
(EMO)
On the
operator
table,
near
joysticks
Push the
button down
until it locks
All systems
apart from
computer are
powered
down -
pumps are
stopped,
manipulator
cannot be
moved with
controls.
Monthly
6.1.2
External connections
Check the routing of all external cables to avoid any physical stress:
•
Install the electrical supply cable outside of usual walkways, to minimise the hazard for
pedestrians.
•
Protect the electrical supply cable from being mechanically stressed (squeezed, pinched,
pulled).
•
Keep the electrical supply cable easily accessible for an emergency disconnection.
•
Protect the electrical supply cable and its terminals from being splashed or immersed in
water.
Do not apply power to the system before having checked that all
cables and pipes are properly connected.
6.1.3
Status indicator LED
There is a coloured indicator LED located on the X-ray source.
The possible colours and their meaning are:
Bright Green
Normal operation
Red (upper position)
“X-ray on” lamp failure
Red (lower position)
“Pre-warning” lamp failure
Orange
Electronics over temperature
Yellow
Oil over pressure
Blue
Coolant flow failure
Содержание XT V 160
Страница 1: ...X Tek X ray and CT Inspection XT V 160 X ray System Operator Manual XTM0011 F2 ...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 8: ......
Страница 22: ......
Страница 26: ......
Страница 32: ......
Страница 83: ...Maintenance XT V 160 X ray System XTM0011 F2 73 ...
Страница 90: ...Troubleshooting 80 XTM0011 F2 XT V 160 X ray System ...
Страница 93: ......