background image

 

 

 

Principles of Modulation Contrast 

 
Since human eyes, cameras, and films capture objects by perceiving differences in light intensity and/or color, they cannot see 
(or sense) the colorless, transparent cells or bacteria. These colorless transparent objects are called “phase objects” since they 
only change the phase of the light when the light passes through them. 
The phase objects can be made visible by dyes, but their life will be deprived. In order to observe the living phase objects, 
differential interference contrast and phase contrast microscopy are invented, and modulation contrast microscopy likewise. The 
modulation contrast microscopy adopts the same optical system as the ordinary microscopes, but with some additional parts 
that convert the transparent specimen into the variation of light intensity. These additional parts modulate the amplitude of the 
light that passed through the specimen, thus changing the intensity of the light making up the visible image. (In differential 
interference contrast and phase contrast microscopy, the phase objects are converted into the variations of light intensity by the 
phase change that occurs when the light passes through the specimen.) 

 

Let’s now think of a light that passes through a specimen. Since a phase object 
has the different refractive index than that of its surroundings, the light will be 
refracted at its border. (See Figure 7 showing refracted light on trapezoidal phase 
object.) The same thing happens to every specimen. 
 
See Figure 8 for the principle of the modulation contrast. 
There is a slit diaphragm on the condenser aperture, and modulator inside the 
NAMC objective. (The modulator is a density filter placed at the exit pupil of the 
NAMC objective. It divides the exit pupil into three regions, dark, half-dark and 
transparent.) 
If there is nothing on the specimen surface, the light passes through the half-dark 
region of the modulator and appears half-dark. If the light is refracted by the phase 
object, it passes either the dark or the bright region according to the difference in 
the refracted angle. The light then appears dark or bright according to the region 
the light passed through. In this way, the phase object is made visible. 
 
In modulation contrast microscopy, the image appears in relief just like the 
differential interference contrast microscopy, though its resolution may be a little 
inferior. The notable point is that there is no influence of double refraction, thus 
enabling you to observe the specimen without double refraction. 

 
 
 
 

Troubleshooting 

 

Problem 

Cause 

Countermeasure 

No polarizing filter attached. 

Screw in the polarizing filter to the condenser mount. 

No NAMC objectives in the optical path. 

Use objectives that have “NAMC” indication on its body. 

NAMC codes on the NAMC objective and the 
NAMC module in the optical path are not the 
same. 

The NAMC objective and the NAMC module in the optical 
path must have the same NAMC code. 

Slit aperture image is not correctly aligned to 
the gray region on the modulator. 

Readjust as described in “Microscopy.” 

Poor or dull 
contrast 

Incorrect condenser position. 

Set the system condenser firmly on the condenser mount. 
Focus the aperture image on the specimen surface and center 
it in the viewfield. 

Unnecessary filters in the optical path. 

Remove unwanted filters (ND filters, etc.) from the optical 
path. 

The DIC analyzer is in the optical path (when 
using the DIC attachment at the same time). 

Remove the analyzer from the optical path. 

Dark Image 

Low illumination voltage. 

Adjust the light intensity control dial to raise the illumination 
voltage. 

Refer to the instruction manual supplied with the microscope for other troubles than mentioned above. 

Figure 7

 

Figure 8

 

Phase object 

Slit aperture diaphragm 

Condenser lens 

Specimen 

Objective 

Modulator 
 

Image

 

Dark Bright 

Содержание NAMC

Страница 1: ...elements made of resin To clean the NAMC module covered with dust or the like lightly blow it off with an air blower Do not use lens cleaning tissue that could damage the optical elements Also avoid using solvents that may damage the optical elements Make sure to keep any dirt from coming into contact with the optical elements during assembly and especially never to touch them with bare hands Asse...

Страница 2: ...he direction shown by the arrow contrast direction Rotate the modulator ring to rotate the modulator so that the specimen can be observed in good condition Be careful not to rotate the correction ring instead of the modulator ring The view of the slit aperture image Observe the two apertures 1 and 2 on the slit aperture image the bridge shaped area 3 is the clearance space between them Note that t...

Страница 3: ...ion Also you can set the clamp ring in the desired position For the Achromat objective however the rotation is limited while no clamp ring is provided During the microscopy note on the following points The NAMC objective and the NAMC module in the optical path must have the same NAMC code Table 1 Combination of NAMC objectives and NAMC modules NAMC Objective NAMC Module NAMC 10x NAMC 1 NAMC 1 NAMC...

Страница 4: ...nt If there is nothing on the specimen surface the light passes through the half dark region of the modulator and appears half dark If the light is refracted by the phase object it passes either the dark or the bright region according to the difference in the refracted angle The light then appears dark or bright according to the region the light passed through In this way the phase object is made ...

Отзывы: