FRAP on the Fly Calibration
Before you can use FRAP the laser must be calibrated
•
Find an empty piece of glass without cells/fluorescence
•
Set the exposure time around 100 - 200ms
•
Open ILAS2 with icon
•
Open calibration tab and put the power of selected laser to 100% (in ILAS2 window)
•
Activate the laser [G] and press “ Show Live”
•
Place laser in upper left corner by moving the red cross, press [H]
•
Repeat for lower right corner
o
If the spot isn’t visible or too big and bright change power and/or exposure time
o
view is rotated, so move laser in other direction
o
keep in mind to move red cross into left and right corners of the calibration screen
•
Calibrate with button [i]
•
To check if your calibration was successful open “ on the fly” tab
•
Change time to 3000
•
Increase the power of your laser and press “ Show Live”
•
Click in the image and check if laser pulse in at the position of your mouse pointer
Troubleshooting - FRAP
FRAP calibration error
•
Notification: Calibration canceled (exceeded 15 bits)
Possible errors
•
Emitting fluorophore or auto fluorescence
o
Ensure that the area of the slide really empty
•
Too much laser power or exposure time
o
Reduce laser power in the ILAS2 window and/or exposure time in Acquire window