Manual: TIRF Microscope

 

Ee-1454

 

 

Martijn de Gruiter 

 

Erasmus OIC 

9

 

 

FRAP on the Fly Calibration 

Before you can use FRAP the laser must be calibrated 

 

 

 

 

 

 

 

 

 

 

 

 

 

Find an empty piece of glass without cells/fluorescence 

 

Set the exposure time around 100 - 200ms 

 

Open ILAS2 with icon 

 

Open calibration tab and put the power of selected laser to 100% (in ILAS2 window) 

 

Activate the laser [G] and press “ Show Live”  

 

Place laser in upper left corner by moving the red cross, press [H]   

 

Repeat for lower right corner 

o

 

If the spot isn’t visible or too big and bright change power and/or exposure time 

o

 

view is rotated, so move laser in other direction 

o

 

keep in mind to move red cross into left and right corners of the calibration screen 

 

Calibrate with button [i] 

 

 

To check if your calibration was successful open “ on the fly” tab 

 

Change time to 3000 

 

Increase the power of your laser and press “ Show Live” 

 

Click in the image and check if laser pulse in at the position of your mouse pointer 

 

 

Troubleshooting - FRAP 

 

FRAP calibration error 

 

Notification: Calibration canceled (exceeded 15 bits) 

Possible errors 

 

Emitting fluorophore or auto fluorescence 

o

 

Ensure that the area of the slide really empty 

 

Too much laser power or exposure time 

o

 

Reduce laser power in the ILAS2 window and/or exposure time in Acquire window  

 

G

i

H

K

J

7

8

Содержание Eclipse Ti TIFT 1454

Страница 1: ...tally reflect on glass medium interface and creates an evanescent wave above the glass Thickness of wave is around 200nm but decays exponentially Possible to only excite fluorophores in membrane without background signal of the cytoplasm Benefits Extreme low background signal Perfect for membrane studies CCD Camera attached will provide in fast imaging Contact Martijn de Gruiter Be 343 tel 31105 G...

Страница 2: ...Manual TIRF Microscope Ee 1454 Martijn de Gruiter Erasmus OIC 2 ...

Страница 3: ...ells uses water o fill reservoir with demineralized water up to level of CO2 tube Allow the system to warm up for 15 30 minutes Settings for the heating unit are paper on the wall choose your settings o General settings for cells on a coverslip o top heater 40 5 C o bath heater 38 5 C o stage heater 38 5 C o lens heater 37 C Ensure that the cells are grown on a glass coverslip Place the slide in t...

Страница 4: ... be done with the mercury lamp o Choose the illumination settings for your fluorophore in the Taskbar in Metamorph o Increase or decrease brightness by changing the ND filters o ND 4 8 10 will block intensity the can be combined use minimal intensity to reduce bleaching of your fluorophore When your sample is in focus you can focus the beam of the laser if necessary for detailed explanation ask a ...

Страница 5: ...ra o right side mercury lamp or transmission to eyepiece Open FRAP or laser console Metamorph Acquire window Sidebar left Exposure time o Minimum interval time at full chip is 90ms quad chip is 50ms o Exposure time can be lower than interval time Full Chip Quad Chip Use Active Region o Use full chip middle quadrant or region selected by ROI Live bin bin o Combine pixels to increase readout speed o...

Страница 6: ...he Illumination choice to Current shutter Tab Special Gain options 10 MHz EM gain gain3 3x is standard setting Control gain between 0 and 1000 Use Frames To Avg option to average images or use longer exposure time Live image window Magnifying glass Display image in other LUT o Grayscale Color by wavelength Histogram of intensities o Manual scaling can be done with the orange arrows Auto scaling op...

Страница 7: ...mes to skip during acquisition o Don t show recording faster when exposure time minimal interval time Camera readout default options should be o Camera state HALT o Shutter mode OPEN PRE SEQUENCE o Clear mode CLEAR PRE SEQUENCE Multi Dimensional Acquisition MDA Combine multiple tasks z stack time lapse streaming multiple positions multiple wavelengths use of journals Open the MDA window using Task...

Страница 8: ...per o Hold lens paper tight between your fingers and wipe over objective o Clean lens with 2 propanol on your lens paper Cover stage with lens paper box Save data to OIC network storage O drive o To collect the data from your own pc connect to network storage drive via the address oic station oic user guest password guest o When the hard disk of the microscopes pc gets full data will be deleted o ...

Страница 9: ...r and or exposure time o view is rotated so move laser in other direction o keep in mind to move red cross into left and right corners of the calibration screen Calibrate with button i To check if your calibration was successful open on the fly tab Change time to 3000 Increase the power of your laser and press Show Live Click in the image and check if laser pulse in at the position of your mouse p...

Отзывы: