![Nikon Eclipse LV100POL Скачать руководство пользователя страница 64](http://html1.mh-extra.com/html/nikon/eclipse-lv100pol/eclipse-lv100pol_instructions-manual_1691841064.webp)
62
Specifications
Model
Optical system
Illumination system
Focusing mechanism
Eyepiece
Stage
Nosepiece
Input ratings
Power cord
Operating conditions
Storage conditions
ECLIPSE LV100POL
CFI60 system (chromatic aberration free infinity system)
Epi/dia selector type
Episcopic illumination:
Built-in type lamp power supply
Specified illuminator:
LV-UEPI Universal Epi Illuminator (NCB11, ND4, and ND16
are installed and can be replaced by users)
Diascopic illumination:
Built-in type lamp power supply, fly's eye lens, NCB11 and
ND8 are installed but cannot be replaced by users
Lamp ratings:
12 VDC, 50 W halogen lamp
Specified lamp:
LV-HL50W, 12 V, 50 W longlife halogen lamp
Specified lamp house:
LV-LH50PC precentered lamp house
Manual operation type single axis coarse/fine focus knob mechanism (calibration marking
for fine focus: 1
µ
m/marking)
Stroke:
30 mm with coarse focus stopper mechanism
Coarse focus knob:
14 mm/revolution
Fine focus knob:
0.1 mm/revolution
10x, field number: 22
Circular graduated stage
Provided with two verniers
Provided with a rotation mechanism with a clamp
Provided with 45
°
click-stop mechanism
Provided with two specimen holders
Quintuple type (five sockets)
Provided with objective centering mechanism
Input voltage:
100-240 VAC
±
10% 50/60 Hz
Rated current:
1.2 A maximum
When the supply voltage is 100 V to 120 V
UL Listed detachable cord set, 3 conductor grounding Type
(3 conductor grounding Type SVT, No.18 AWG, 3m long
maximum, rated at 125V AC minimum)
When the supply voltage is 220 V to 240 V
Approved according to EU/EN standards, 3 conductor
grounding Type (3 conductor grounding Type H05VV-F, 3m
long maximum, rated at 250V AC minimum)
Temperature:
0
°
C to +40
°
C
Humidity:
85% relative humidity maximum (no condensation)
Altitude:
2000 m maximum
Pollution degree:
Degree 2
Installation category:
Category II
Electric shock protection class: Class 1
Indoor use only
Temperature:
-20
°
C to +60
°
C
Humidity:
90% relative humidity maximum (no condensation)