WTI-8038RL 

WTI-8038RL 

90°

90°

For Best accuracy, use a reference master the same 

height as the workpiece. 

 Attach Mounting Stem.

Loosen the Stem mounting bracket and slide the groove 

onto  the  Dovetail  in  the  desired  position.  Tighten  the 

Stem to secure. (Choose a stem with diameter to match 

the Gauge Holder you plan to use.)

Confirm that the Probe and stem are secure 

without any play.

If they are loose, tighten to secure.
Attach the Indicator Gauge to the Holder. 
Please attach by the Stem only. 
Confirm that the pointer moves freely.

Bring the Probe into contact with the workpiece, 

and gently raise and lower the gauge while 

checking the movement. Confirm that the pointer 

returns to initial position when raised.
Make sure the Probe is parallel to the measured 

surface, and perpendicular to the direction of 

the measurement. 
If probe is not positioned properly, the measurement 

will be inaccurate and operation may be faulty.

HOW TO USE - Preparation

HOW TO USE - Comparison Measurements

Set Up Reference Part

Position the reference Part with the measured surface 

parallel to the probe and perpendicular to the direction of 

the measurement. Bring into contact with the probe, but 

use care that the part does not bump the probe from the 

side.
Set the Origin.

Adjust the gauge mount or rotate the Bezel to set 

the Gauge to “0”.
Remove the Reference Part and Measure the 

Workpiece

Remove the master with care not to jostle the gauge. 

Replace with the workpiece and read the scale. 

How to Use - (Parallelism, Flatness, Runout, etc.)

Position the probe on the workpiece.
Position the Part with the measured surface parallel to the probe and perpendicular to the direction of the 

measurement. Bring into contact with the probe, but use care that the part does not bump the probe from the side.
Set the Origin.
Adjust the gauge mount or rotate the Bezel to set the Gauge to “0”.
Move the workpiece and read the scale.
Move the work slowly while reading the position of the pointer on the scale.

 Rotate to remove

Dovetail

Insert

Insert

Insert

Dovetail

Groove

Stem Mount

Stem

φ

6・

φ

8mm)

Ex.

Ref. Part 30mm Gauge Block

:30.00mm

Workpiece height

:30.08mm

Set Origin with Reference Part

〉 〈

Measure

 Pointer

Scale 

     Plate

 Bezel

Dir. of Measurement

Dir. of Measurement

Dir. of Measurement

Dir. of Measurement

Slightly Loose

Tighten