background image

                

CGT1050 

 GT1060

 

CO

2

/TEMP/RH METER

                

_____________________________________________________________________________________________

 

 

INTRODUCTION  

Thank you for purchasing this portable CO

2

 meter. The meter measures CO

level, air temp., 

dew point, wet bulb temp. and humidity(DP, WB, RH are for model GT1060-RH) and is an ideal 
instrument for indoor air quality (IAQ) diagnosis.  
 
Poor indoor air quality is considered unhealthy because it causes tiredness, loss of ability to 
concentrate, and even illness(ex. Sick Building Syndrome). IAQ monitoring and survey, 
especially on CO2 level and air ventilation become widely applied in public areas such as 
offices, classrooms, factories, hospitals and hotels. It is also suggested in regulations of 
industrial hygiene in some countries. (Appendix). 
 
The portable CO2 meter uses NDIR (non-dispersive infrared) technology to ensure the 

reliability and long term stability. It’s useful in verifying HVAC system performance and air 

ventilation control. 
 

Features:  

 

Triple displays of CO

2

 level, temp. and humidity (GT1060 RH).  

 

Stable NDIR sensor for CO

2

 detection 

 

Statistics of weighted averages 
  TWA 

(8 hours weighted average)

 

  STEL 

(15 minutes weighted average)

  

 

Backlight for working in dark area  

 

Audile CO

2

 warning alarm  

 

Battery and adaptor power supply  

 

Easy manual calibration on CO

and humidity  (RH for GT1060 RH only) 

 

MATERIAL SUPPLIED  

 

Instrument 

 

4pcs AA batteries 

 

Operation manual 

 

Hard carrying case  

 

POWER SUPPLY 

The meter is powered by either 4 AA batteries or a DC adaptor(9V/1A output).  
 
Install the batteries into the battery compartment on the rear and make sure they are in 
correct polarity and good contact. When an adaptor is used, it will cut off the power supply 

from batteries. The adaptor can’t be used as a battery charger. 

 

 
When battery voltage gets low, 

 

and “Lob” will appear on the LCD (Fig.1). And beeper sounds. 

The CO

2

 

sensor can’t work under low voltage, so it beeps to indicate failed CO

2

 measurement  

(press any key but 

 

to stop the beeps) and the readings won’t be displayed. Please 

replace with fresh batteries or connect with an adaptor.  
  
 
 
 
 

 

 

 

    GT1050                             GT1060 

Содержание GT1050

Страница 1: ...USER MANUAL GT1050 GT1060 CO2 TEMP RH METER ...

Страница 2: ...ance and air ventilation control Features Triple displays of CO2 level temp and humidity GT1060 RH Stable NDIR sensor for CO2 detection Statistics of weighted averages TWA 8 hours weighted average STEL 15 minutes weighted average Backlight for working in dark area Audile CO2 warning alarm Battery and adaptor power supply Easy manual calibration on CO2 and humidity RH for GT1060 RH only MATERIAL SU...

Страница 3: ...GT1060 RH AIR Air temperature WBT Wet bulb temperature GT1060 RH Unit of relative humidity oE C F Celsius Fahrenheit KEYPAD Turns on and off the meter Enters setup mode Sets as non sleep mode with Exits setup page mode Enters CO2 calibration with Enters RH calibration with Freezes the current readings Cancels data hold function Activates or cancels the backlight Selects unit or increases value in ...

Страница 4: ... measurement when power on and update readings every second In the condition of operating environment change ex from high to low temp it takes 30 sec to respond for CO2 sensor and 30 minutes for RH NOTE Do not hold the meter close to faces in case exhalation affects CO2 levels AIR all models DP WBT GT1060 RH Press to switch temperatures display The lower left display will cycle from air temperatur...

Страница 5: ...r WB temperatures and humidity GT1060 RH on the lower displays Fig 7 In STEL and TWA modes the main display shows the weighted average of CO2 readings for the past 15 minutes STEL and 8 hours TWA The lower displays are the current AIR DP WB temperatures and humidity GT1060 RH Fig 8 NOTE 1 If the meter is turned on for shorter than 15 minutes the STEL value will be the weighted average of readings ...

Страница 6: ... be disabled during calibration mode SETUP Hold down under normal mode for more than 1 sec to enter setup mode To exit setup press in P1 0 or P3 0 and it returns to normal mode Note P2 0 is not applicable in these models but for future model with CO and CO2 measurement When entering setup mode P1 0 and AL Fig 10 are displayed on the LCD Press to go into P1 1 for setting CO2 alarm threshold The cur...

Страница 7: ... calibration site Turn on the meter and hold down and simultaneously to enter CO2 calibration mode Fig 14 400ppm and CAL are blinking on the LCD while performing calibration Wait about 10 minutes until the blinking stops and the calibration is completed automatically and back to normal mode To abort the calibration turn off the meter at any time NOTE Ensure the batteries are with full voltage duri...

Страница 8: ...60 minutes when CAL and humidity stop blinking Fig 16 After 33 calibration plug the sensor probe into 75 salt bottle then press to enter 75 calibration Fig 17 CAL and calibrating value 75 2 if at 25 C are blinking on the LCD with current temperature at the left Instrument is now calibrating Wait about 60 minutes until blinking stops then calibration is completed and it returns to normal mode NOTE ...

Страница 9: ...1 CO2 sensor damaged E02 The value is under range E03 The value is over range E04 The original data error results in this error DP WB E07 Too low voltage to measure CO2 Replace batteries or use an adaptor E11 Retry humidity calibration E17 Retry CO2 calibration E31 Temperature sensor damaged E34 Humidity sensor damaged PC CONNECTION The meter can do PC link for on line logging and data analysis vi...

Страница 10: ...CGT1050 GT1060 CO2 TEMP RH METER _____________________________________________________________________________________________ SPECIFICATION Model GT1050 Model GT1060 ...

Страница 11: ...ion is 30 000ppm ASHARE Standard 62 1989 Sec 6 1 3 Comfort odor criteria are likely to be satisfied if the ventilation rate is set so that 1 000 ppm of CO2 is not exceeded WARRANTY The instrument is warranted to be free from defects in material and workmanship for a period of one year from the date of purchase This warranty covers normal operation and does not cover misuse abuse alteration neglect...

Страница 12: ......

Отзывы: