11

 

 

 
8.   If the 2

nd 

 point calibration is required then rinse the electrode with distilled water 

      and immerse in pH4.01 or pH10.01, wait for stability 

, otherwise press 

ESC

 to finish  

      and exit. 
9.   Repeat the points 4 to 7. 
10.  If the 3

rd

 point calibration is required then rinse the electrode with distilled water and 

      immerse in the last buffer, wait for stability 

, otherwise press 

ESC

 to finish and exit. 

11.   Repeat the points 4 to 7. 
12.  De calibration procedure is completed and the meter is ready to use. 

 

NOTE: Anytime press 

ESC

 key to abort and exit from calibration procedure. 

 

POWER OFF 

 

To turn off the meter press       key for 3 seconds, the meter will switch off. The instrument can not be 
switched off during calibration. 
 

REPLACEMENT OF SENSOR 

 

Tester 5 series has replacement sensors which can be replaced in case it expires of damages. 
 
1) To replace the sensor unscrew the dial in anti-clock wise. 
2) Pull out sensor from unit body. 
3) Put a new sensor by matching correctly the sign of dent. 
4) Be sure that all the gaskets are good and in correct position. 
5) Screw the dial tightly. 
 

        SENSOR MAINTENANCE 

 

Before first time use or after long time dry storage, leave the probe in tap water or storage solution 
for at least 30 minutes to activate the sensor. 

 

EC2620-Cond1 / EC2630-Cond5: 
- Rinse the probe with distiller water or sample before use. 
- For short time storage use distiller water. For long time store dry. 

 

PH2620-pH1 / PH2630-pH5 pH5 Food / PH2720-ORP5 / EP2000-PC5:  
- Rinse the probe with distiller water or sample before use. 
- Store the probe in storage solution for regular use. For long time store dry. 
- Never store pH sensor in distilled water 
 
 

Never touch conductivity probe with paper, for cleaning purpose only rinse with distilled water. 
If touched the probe may damage. 

 

 
 
 

 

Содержание EC2620

Страница 1: ...USER MANUAL Testers EC2620 EC2630 EP2000 PH2620 PH2630 PH2720...

Страница 2: ...PLY 6 INSTRUCTIONS FOR PH2620 pH1 EC2620 Cond1 7 POWER ON 7 POWER OFF 7 SETUP MENU for PH2620 pH1 EC2620 Cond1 7 MEASUREMENT 7 CALIBRATION PROCEDURE FOR CONDUCTIVITY EC2620 Cond1 8 CALIBRATION PROCEDU...

Страница 3: ...TESTERS EC2620 EC2630 EP2000 PH2620 PH2630 PH2720 _____________________________________________________________________________________________ 3 PRODUCT SPECIFICATIONS...

Страница 4: ...3 key for all functions SAFETY INSTRUCTIONS Read this instruction manual carefully before using your new tester The membrane of pH electrode is made of glass and can be danger in case it breaks To av...

Страница 5: ...______________________________________________________________________ 5 PRODUCT DESCRIPTION KEYPAD PH2630 pH5 EC2630 Cond5 EP2000 PC5 PH2620 pH1 EC2620 Cond1 Keypad Functions for Tester PH2630 pH5 EC...

Страница 6: ...CALIBRATION POINTS INDICATOR When a calibration is performed these icons indicate the points calibrated and First point for pH calibration is always 7 00 pH POWER SUPPLY This tester series works with...

Страница 7: ...o select the parameter between COND or TDS to be used for measurements and press to confirm only for EC2620 Cond1 o Only if the TDS is selected then the display will show TDS FCT flashing press to cha...

Страница 8: ...d solution CALIBRATION PROCEDURE FOR TDS EC2620 Cond1 When the meter is set to read TDS then the calibration is done on TDS with 1 or 2 points The calibration procedure for TDS is same as for Conducti...

Страница 9: ...2630 Cond5 EP2000 PC5 1 With meter switched off press and hold CAL key and press key once 2 The meter will switch on with all the segments active release CAL key the meter will got into the Setup Menu...

Страница 10: ...tandard value will flash for 3 times and then the meter will go into the measurement mode 8 If the 2nd point calibration is required then rinse the electrode with distilled water and immerse in the 2n...

Страница 11: ...bration REPLACEMENT OF SENSOR Tester 5 series has replacement sensors which can be replaced in case it expires of damages 1 To replace the sensor unscrew the dial in anti clock wise 2 Pull out sensor...

Страница 12: ...suring value is not stable during calibration Press key when icon appears During calibration the measuring value is not stable for 3min 1 Check whether there are bubbles in glass bulb 2 Replace with n...

Страница 13: ......

Отзывы: