background image

                

TESTERS

 

EC2620 / EC2630 / EP2000 / PH2620 / PH2630 / PH2720

                

_____________________________________________________________________________________________

 

10

 

 

 

MEASUREMENT 

 

Rinse the electrode with distilled water or sample prior to start measurement. 
Fill the measuring cap with sample, switch on the meter with      and press 

MODE

 key to select 

desired parameter to be measured (green backlight during measurement). 
Immerse the Tester in sample and wait for stability, when stability icon 

 appears on display take 

the reading. 

During measurement make sure that pH electrode is free from air bubbles, and that there isn’t 

any air bubble around or between conductivity sensor. 

 

CALIBRATION PROCEDURE FOR CONDUCTIVITY (EC2630-Cond5/EP2000-PC5) 

 

1.    Power ON the meter by pressing the       key. 
2.   Rinse the probe with distilled water.

 

 

3.   Immerse the probe in the calibration solution (84

μ

s, 1413

μ

S or 12.88 mS), wait for 

      stability 

  

4.   Press 

CAL

 key, (blue backlight during calibration). 

5.   The meter will start calibration procedure and will recognise automatically the 
      standard used. 
6.   When stable press          to confirm and complete the calibration. 
7.    The standard value will flash for 3 times and then the meter will go into the 
       measurement mode. 
8.   If the 2

nd

 point calibration is required then rinse the electrode with distilled water  

      and immerse in the 2

nd

 Standard, wait for stability 

  

9.   Repeat the points 4 to 7. 
10.  If the 3

rd

 point calibration is required then rinse the electrode with distilled water 

      and immerse in the 3rd Standard, wait for stability 

  

11.   Repeat the points 4 to 7. 
12.  The calibration process is completed and the meter is ready to use. 
 
NOTE: Anytime press 

ESC

 key to abort and exit from calibration procedure. 

NOTE(2): In case multipoint calibration is performed it is better to start from the lower value 
standard first and then go increasing. 

 

CALIBRATION PROCEDURE FOR pH (PH2630-pH5, EP2000-PC5) 

 

1.    Power ON the meter by pressing the        key. 
2.   Rinse the electrode with distilled water.

 

 

3.   Immerse the electrode in the 1

st

 buffer solution pH7.00 and wait for stability 

  

4.   Press 

CAL

 key (blue backlight during calibration). 

5.   The meter will start calibration procedure and will recognize automatically the 
      standard used. 
6.   When stable 

 press          to confirm and complete the calibration. 

7.   The standard value will flash for 3 times and then the meter will ask for next point 
      for calibration. If only 1 point calibration is required then press       to finish and exit. 
 
 

 

Содержание EC2620

Страница 1: ...USER MANUAL Testers EC2620 EC2630 EP2000 PH2620 PH2630 PH2720...

Страница 2: ...PLY 6 INSTRUCTIONS FOR PH2620 pH1 EC2620 Cond1 7 POWER ON 7 POWER OFF 7 SETUP MENU for PH2620 pH1 EC2620 Cond1 7 MEASUREMENT 7 CALIBRATION PROCEDURE FOR CONDUCTIVITY EC2620 Cond1 8 CALIBRATION PROCEDU...

Страница 3: ...TESTERS EC2620 EC2630 EP2000 PH2620 PH2630 PH2720 _____________________________________________________________________________________________ 3 PRODUCT SPECIFICATIONS...

Страница 4: ...3 key for all functions SAFETY INSTRUCTIONS Read this instruction manual carefully before using your new tester The membrane of pH electrode is made of glass and can be danger in case it breaks To av...

Страница 5: ...______________________________________________________________________ 5 PRODUCT DESCRIPTION KEYPAD PH2630 pH5 EC2630 Cond5 EP2000 PC5 PH2620 pH1 EC2620 Cond1 Keypad Functions for Tester PH2630 pH5 EC...

Страница 6: ...CALIBRATION POINTS INDICATOR When a calibration is performed these icons indicate the points calibrated and First point for pH calibration is always 7 00 pH POWER SUPPLY This tester series works with...

Страница 7: ...o select the parameter between COND or TDS to be used for measurements and press to confirm only for EC2620 Cond1 o Only if the TDS is selected then the display will show TDS FCT flashing press to cha...

Страница 8: ...d solution CALIBRATION PROCEDURE FOR TDS EC2620 Cond1 When the meter is set to read TDS then the calibration is done on TDS with 1 or 2 points The calibration procedure for TDS is same as for Conducti...

Страница 9: ...2630 Cond5 EP2000 PC5 1 With meter switched off press and hold CAL key and press key once 2 The meter will switch on with all the segments active release CAL key the meter will got into the Setup Menu...

Страница 10: ...tandard value will flash for 3 times and then the meter will go into the measurement mode 8 If the 2nd point calibration is required then rinse the electrode with distilled water and immerse in the 2n...

Страница 11: ...bration REPLACEMENT OF SENSOR Tester 5 series has replacement sensors which can be replaced in case it expires of damages 1 To replace the sensor unscrew the dial in anti clock wise 2 Pull out sensor...

Страница 12: ...suring value is not stable during calibration Press key when icon appears During calibration the measuring value is not stable for 3min 1 Check whether there are bubbles in glass bulb 2 Replace with n...

Страница 13: ......

Отзывы: