TEST METHOD
Binocular Visual Function Test
4.4.3
Phoria test
Use: To detect exophoria, esophoria, hypophoria, and hyperphoria
Chart: Phoria
Auxiliary lens: Right eye 135°, left eye 45°, polarizing filters, binocular rotary prism
(SC-2000, SC-1600: Red filter on the right eye, green filter on the left eye, binocular rotary prism)
Type U
Ideal appearance:
Right eye
Left eye
Binocular vision (Orthophoria)
[Procedure example]
1
Present the phoria chart.
Press
.
The polarizing filters are placed in the measuring windows.
The base in/out prism powers can be entered.
When
is pressed again, the base up/down prism powers can be entered.
2
Ask the patient, “Can you see the four bars?”
Yes
Go to the next step.
No
The test cannot be continued.
3
Ask the patient, “Do the vertical and horizontal bars form a cross?”
Yes
Orthophoria
No
Heterophoria - Go to the next step.
4
Ask the patient, “Do the vertical bars line up?”
Yes
No esophoria nor exophoria
No
Esophoria or exophoria
5
Ask the patient, “Is the top bar shifted to the left or right?”
Right
Esophoria
Left
Exophoria
6
Ask the patient, “Do the horizontal bars line up?”
Yes
No hyperphoria nor hypophoria
No
Hyperphoria nor hypophoria
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...