5 - 8
9. Make a clearance of 0.5 mm between the mask and chart disks by displacing the chart
ASSY. Tighten HH4 × 6 (n=4) of the chart disk retainer.
* Tighten the four screws (HH4 × 6(n=4)) while checking that the two tips contact the notch
of the motor shaft (flat side) vertically.
* Check that the mask and chart disks do not contact each other by rotating them by hand.
(As for the bottom of the disk housing, check the disks by shining light in there with a
penlight.)
10. Assemble the removed parts (except for the bottom and top cover ASSYs) in reverse
order.
* Check that the shading plate of the chart disk does not interfere with the chart sensor.
* Write down the connector number “P5” on the connector of the replaced mask motor with
an oil-based ink (black).
* Tie the wiring with the tie wrap (T18R) in its original position. (See “7.2 How to carry out
the wiring”.)
11. After performing “6.1 Adjustment of the initialized position”, “6.2 Adjustment of the
sensor voltage” and “6.4 Adjustment the fixed mask position” and “6.5 Adjustment of
the chart and mask positions”, remount the bottom and top cover ASSYs.