5 - 6
5.3.2 Replacement of the mask sensor
Replacement part: Mask sensor (33155-EA05)
1. Remove the top cover ASSY. (See 5.1.)
2. Disconnect the connector from J4 on the BA01 board.
* Cut the tie wrap (T18R) which ties the cables of the mask sensor.
3. Remove the disk cover (M030). (See 5.1.)
4. Unscrew BS3 × 6
ZnC
(n=2)
to remove the
mask sensor and sensor fitting plate
(M009).
5. Unscrew PC3 × 6 (n=2) from the sensor
fitting plate (M009) and replace the mask
sensor (EA05) with a new one.
* Check that the shading plate of the mask
disk does not interfere with the mask
sensor.
6. Assemble the removed parts (except for the top cover) in reverse order.
* Write down the connector number “P4” on the connector of the replaced mask sensor with
an oil-based ink (black).
* Tie the wiring with the tie wrap (T18R) in its original position. (See “7.2 How to carry out
the wiring”.)
7. After performing “6.2 Adjustment of the sensor voltage”, remount the top cover.
ZnC
BS3×6
センサー取付板
PC3×6
2
2
マスクセンサー
Sensor fitting plate
Mask sensor
PC3×6
2
BS3×6
ZnC
2
Содержание CP-690
Страница 1: ...MCP7 RDA001B E Total Page 70 2006 7 NIDEK AUTOMATIC CHART PROJECTOR Model CP 690 SERVICE MANUAL...
Страница 6: ...1 2...
Страница 8: ...2 2...
Страница 44: ...5 22...
Страница 55: ...INFORMATION 7 7 1 Wiring Diagram Total wiring Diagram EA13 BA14 RESISTOR BOARD 2 2 EA14 CHART LED...
Страница 56: ...7 2 EA01...
Страница 57: ...7 3 EA02...
Страница 58: ...7 4 EA05 EA06 EA13 EA14 BA01 Describe the connector No when assembling Mask side P4 Chart side P6...
Страница 60: ...7 6...
Страница 70: ...9 2...