12
Return to previous page
7 ) After loosening the screws fastened by a threadlocking adhesive, be sure to reapply
the threadlocking adhesive to the screws before retightening them.
8 ) After replacing parts, confirm that they are fastened in their original positions securely
before turning on the power.
9 ) See Wiring Diagram and Connecting Cable to check cable breaks as described in
TROUBLESHOOTING. In addition, check cables for the following:
a . Connectors and contact pins are connected securely.
b . Contact pins are crimped securely.
c . No contact failure occurs after re-connection of connectors.
d . Cables are soldered properly.
* Do not pull on the cables with excessive force. Doing so may result in cable
breakage. Never conduct repairs with wet hands. Electric shock or failure of the
device may result.
10) The
labeled area indicates high voltage. Take proper precautions against electric
shock.
11 ) When a cable with any of the following marks is disconnected and reconnected, be
sure to confirm that the screws are tightened securely.
a . Protective ground:
b . Functional ground:
c . Equipotential:
12) Install the device where the outlet which the power plug is inserted into is easily
accessible during use. In addition, ensure that the power plug can be disconnected
without the use of a tool.
Otherwise, it may interfere with disconnection of the device from the input power
source in case of abnormality
5 . Adjustment precautions
1 ) Confirm that the following conditions are met before installation.
a . A place which is level and stable without vibration and shock
b . No exposure to direct sunlight or ultraviolet rays
c . A place with minimal external light sources
d . No exposure to water
e . A place where temperature and humidity meet the specifications for use
f . Dust-free and smoke-free
2 ) Never use the adjustment jigs for purposes other than those intended.
6 . Caution after repairs
3 ) Confirm that the device operates properly after repairs
Содержание AR-310A
Страница 2: ......
Страница 10: ...XAR17ARDA001G E 10 Return to previous page This page is intentionally left blank ...
Страница 62: ...XAR17ARDA001G E 62 Return to previous page This page is intentionally left blank ...
Страница 68: ...XAR17ARDA001G E 68 Return to previous page This page is intentionally left blank ...
Страница 102: ...XAR17ARDA001G E 102 Return to previous page This page is intentionally left blank ...
Страница 186: ...XAR17ARDA001G E 186 Return to previous page 9 3 1 2 30601 E000 AR 310A and ARK 510A ...
Страница 187: ...XAR17ARDA001G E 187 Return to previous page 9 3 1 3 30603 E000 AR 360A and 30703 E000 ARK 560A ...
Страница 188: ...XAR17ARDA001G E 188 Return to previous page 9 3 2 Partial writing diagram 30601 EA01 30601 EA11 30601 EA12 ...
Страница 189: ...XAR17ARDA001G E 189 Return to previous page 30601 EA13 Cable 1 P301 A row side Cable 2 P301 B row side ...
Страница 190: ...XAR17ARDA001G E 190 Return to previous page 30601 EA14 ...
Страница 192: ...XAR17ARDA001G E 192 Return to previous page 30601 EA23 30601 EA24 ...
Страница 194: ...XAR17ARDA001G E 194 Return to previous page 30601 EA30 30601 EA31 30601 EA32 ...
Страница 195: ...XAR17ARDA001G E 195 Return to previous page 30601 EA34 30601 EA36 30601 EA39 30601 EA42 30601 EA44 ...
Страница 196: ...XAR17ARDA001G E 196 Return to previous page 30601 EA51 30601 EA52 ...
Страница 197: ...XAR17ARDA001G E 197 Return to previous page 30601 EA53 30601 EA54 30601 EA55 Motor is included ...
Страница 198: ...XAR17ARDA001G E 198 Return to previous page 30601 EA63 30601 EA64 30601 EA65 30601 EA72 30601 EA75 ...
Страница 199: ...XAR17ARDA001G E 199 Return to previous page 30601 EA76 30601 EA80 30601 EA81 30601 EA82 30601 EA83 ...
Страница 217: ......