manualshive.com logo in svg
background image

4

Operation Method/Functional Explanation

LIQUID TYPE

MODE

ZERO

SET

 UP

HOLD

POWER

kPa

HOLD

LIQUID TYPE

MODE

ZERO

SET

 UP

HOLD

POWER

kPa

O.  SET

LIQUID TYPE

MODE

ZERO

SET

 UP

HOLD

POWER

kPa

[Insertion of Batteries]

Slide the battery cover on the back of the product body downward to open the cover. Insert the batteries in the direction 
as illustrated, check that the batteries are correctly and firmly in place, and slide the battery cover again to close.

[Operation Method]

(1)Turning the power ON

 With  the  power  turned  off,  press  the    POWER    button  under  open  air  atmospheric  pressure,  and  check  that  the 

minimum digit for the pressure indication is within 0±10 digits.(For the pressure range “102A”, 100(kPa abs.)will 
be displayed when the power button is turned on under atmospheric pressure.)

(2)Zero adjustment

 After turning on the power, if a difference is detected between the zero values in open air atmospheric pressure, 

press  the    ZERO    button  for  three  seconds  or  longer  and  performs  the  zero  adjustment.(The  zero  adjustment 
function is invalid in 102A range, because it is based on an absolute vacuum.) 

(3)Measurement

 When pressure is applied, the measured value is displayed on the LCD.

(4)Turning the power OFF

 With the power turned on, press the  POWER  button to turn off the power. When the auto power off function is 

activated, the power is turned off according to the preset time.

[Functional Explanation]

(1)Hold

 When  pressure  is  applied,  the  measured 

value is displayed. Press the HOLD button 

to  retain  the  pressure  value  displayed  at 

that time. At this time, the “HOLD” mark 

is displayed on the LCD. Press the button 

again to cancel the hold.

(2)Offset

 When  the  pressure  exceeds ±1%FS,  press 

the    ZERO    button  for  three  seconds  or 

longer  to  enable  subsequent  indication  of 

variations  with  the  zero  value  set  to  0.  At 

this time,“0.SET” is displayed on the LCD. 

Press the  ZERO  button for three seconds 

or  longer  again  to  return  the  indication  to 

the  previous  display.(The  offset  function  is 

invalid in 102A range, because it is based on 

an absolute vacuum.)

(3)Switching of auto power off function

 With  the  power  turned  off,  press  the  

POWER  button for one second or longer 

while  pressing  the    HOLD    button  to  set 

the mode to the switching mode of the auto 

power off function. Choose the mode from 

four modes including [“P3”: 3 min.],[“P20”: 

20 min.],[“PAt”: 20 min. after the pressure 

reaches zero], and [“PoF”: Disabled].

 Press  the    ZERO    button  to  select  the 

mode  and  press  the    HOLD    button  to 

enter the mode. While the switching mode 

is activated, if any button operation is not 

made  for  20  seconds,  the  mode  shifts  to 

the measurement mode at the preset value 

displayed at that time. In the initial state, 

[“P3”: 3 min.] is set.

(4)Unit switching function (Only for −W model)

 With  the  power  turned  off,  press  the  

POWER    button  for  one  second  or  longer 

while  pressing  the    HOLD    button  to  set 

the mode to the unit switching mode. While 

the  unit  indication  is  blinking,  press  the  

ZERO  button to select the mode and press 

the  HOLD  button to enter the mode.

(5)Battery replacement timing

 When the battery voltage is low, “LOBAT” 

is  displayed  on  the  LCD.  Replace  the 

batteries. The batteries are two AAA dry 

cells.  The  used  batteries  should  be 

discarded  according  to  the  instructions  of 

the  respective  local  governments.  After 

replacing the batteries, be sure to perform 

the  zero  adjustment.(See  Operation 

Method(2). But excluding 102A range.)

(6)Excessive pressure error

 When  pressure  exceeding  130%  of  the 

rated  pressure  is  applied,  the  indication 

will  start  to  blink.  Check  the  source 

pressure,  and  use  the  product  within  the 

range of the rated pressures.

bar psiin.

LIQUID TYPE

MODE

ZERO

SET

 UP

HOLD

POWER

M kPa

 mmHg Aq

LIQUID TYPE

kPa

LOBAT

R03,LR03

With the power 

turned OFF, 

press  HOLD  and   

POWER  for one 

second or longer.

With the power 

turned OFF, 

press  ZERO  and 

POWER  for one 

second or longer.

《Unit Switching Function》

(Only for −W model)

《Switching of Auto Power OFF Function》

Operation Mode

(1)3 minutes

(2)20 minutes

(3)20 minutes after the

  pressure reaches zero

(4)Disabled

ZERO button: SELECT

HOLD button: ENTER

ZERO button: SELECT

HOLD button: ENTER

(1)kPa
(2)MPa
(3)mmHg
(4)in.Hg

(5)mmAq
(6)psi
(7)bar

Содержание PG-100B Series

Страница 1: ...ntionally in the air by holding the strap may result in unexpected accidents injury or product damage Be sure to use the product with its main body gripped firmly 6 Be careful not to give external shocks to the products External shocks such as hitting and dropping may cause product failure 7 Installation of attachments When installing an attachment to the pressure port the tightening torque should ...

Страница 2: ... be careful not to allow the neutral detergent solution to enter the pressure port and or main body Infiltration of the neutral detergent may cause product failure 6 Caution for storage of protect cover Please store the protect cover in a cool and dark place if it is to be stored Ultraviolet rays exposure may make the quality of the cover deteriorate Please replace the protect cover with new one in...

Страница 3: ...ero adjustment Hold Offset Auto power off Unit switching 0 5 F S 2digits Gases Liquids compatible with SUS316L 0 to 50 No freezing 15 to 85 RH or below No condensing 3 VDC AAA dry cell x 2 M5 female screw P 0 8 IP40 IEC standards 55 W 16 9 D 110 H Approx 100g Approx 1 000 hours 5 0 5V DC 5mA maximum External supply 1 000V 10mV Indicated value standard 102A Absolute 103R 102R 2 0 F S 2digits 1 0 F S ...

Страница 4: ... function is invalid in 102A range because it is based on an absolute vacuum 3 Switching of auto power off function With the power turned off press the POWER button for one second or longer while pressing the HOLD button to set the mode to the switching mode of the auto power off function Choose the mode from four modes including P3 3 min P20 20 min PAt 20 min after the pressure reaches zero and PoF ...

Страница 5: ...function or damage to the main body Please attach the protect cover to the main body from top side the side of pressure port to the end completely Be careful about the front side of the cover Strap canʼt be attached with the protect cover on the main body Periodical Calibration If you would like to calibration request please contact the sales office where you purchased Warranty The warranty period f...

Отзывы: