SPIROMETER SM-1
CLIFTON RANGE® - 1 YEAR WARRANTY 

spirometer and the other to the water tap. Fill the flask to the mark and air will be 

despatched from the flask into the spirometer. Close the clip and empty the flask.

10.  A second line or reading should appear above zero volume line which should provide 

a scale e.g 500ml air volume = height ?mm. This is the basis for defining the scale. 

11.  Set the valve to SPIROMETER and adjust the weight so that the float comes to rest 

in the mid position.

12.  If oxygen consumption is to be measured fill the nylon net bag with coarse soda 

lime and place it in the container (front view, right hand side of the tank, unscrew 

the wing nut and remove the cap).

13.  Ensure there are no gaps between the net bag and the walls of the container. 

Replace the cover and tighten. 

14.  Gently press down on the float to expel the air and turn the valve to ATMOSPHERE.

15.  Connect the medical grade oxygen supply to the oxygen inlet (container, adjacent 

to the wing nut tube) using pressure tubing.

16.  Slowly turn on the oxygen supply. Oxygen will flow into the unit and raise the float. 

When it contains about 2 litres turn off the supply set the valve to SPIROMETER and 

press down on the float. This forces the oxygen out through the mouthpiece and 

flushes the air out of the system. Do not allow oxygen to be discharged near a naked 

flame. 

17.  Repeat the process, filling and flushing and then close off the oxygen supply and 

turn the valve to ATMOSPHERE. If the cylinder is to be removed ensure a tube and 

clip are left closing the oxygen inlet. 

OPERATING INSTRUCTIONS

1.  A two way valve is fitted. When this valve is set to ATMOSPHERE the breath-

ing tubes are open to the atmosphere and the spirometer is closed. The subject 

therefore breathes normal atmospheric air. This position should be used when the 

subjects first use the spirometer to enable them to become accustomed to it.

2.  When the valve is set to SPIROMETER the subject takes in the oxygen contained 

under the float, causing the float to fall. The expired gases pass through the soda 

lime contained before returning to the float. The soda lime absorbs the carbon di-

oxide from the expired gases. The valve is set to this position when the spirometer 

is in normal use. 

 

 

Medical grade oxygen cylinder must only be used for respiration  

 

 

 

purposes. Before using soda lime dust should be removed from it by   

 

 

pouring the soda lime from one container to another in the open air.

  

!

Содержание Clifton Range SM-1

Страница 1: ...High quality products at affordable prices SPIROMETER SM 1...

Страница 2: ......

Страница 3: ...should only be used under the direct supervision of a teacher who is experienced in their correct use Teachers are reminded that they have a special responsbility when pupils are used as the subject o...

Страница 4: ...designed to offer minimum resistance to air flow This means greater comfort for the user and more realistic results The spirometer enables rate of breathing volumes of breaths and consumption of oxy...

Страница 5: ...anywhere in the system as oxygen can form explosive compounds with these materials 4 Immerse the mouth piece into sterilizing solution Spare mouth pieces are available from your distributor BX0612 Dis...

Страница 6: ...res turn off the supply set the valve to SPIROMETER and press down on the float This forces the oxygen out through the mouthpiece and flushes the air out of the system Do not allow oxygen to be discha...

Страница 7: ...by the amplitude of the trace 4 Respiratory rate the number of breathes per minute or more formally the number of movements indicative of inspiration and expiration is usually determined by counting...

Страница 8: ...new condition Descale by adding 1 litre of vinegar to water and leave for an hour Empty and brush the lime away Rinse thoroughly afterwards The stainless steel crevice free tank should provide years o...

Страница 9: ...y materials or workmanship in any goods manufactured supplied by Nickel Electro Ltd which under proper and normal conditions of use may develop within a period of one years from the date of delivery 3...

Страница 10: ...for specialist engineers currently in force 7 In the event of replacement with a new or reconditioned model the replacement unit will continue the warranty period of the original equipment 8 If any g...

Страница 11: ...rbalance weight is supplied Slide the weight onto the rod and tighten the thumb screw This is the counterweight assembly for the float 2 The large white moulding is the float assembly a metal frame is...

Страница 12: ...rew is fitted Each one locates into a dimple on the top rim of the tank The float assembly will see saw adjust the counter balance weight accordingly 5 In a polythene bag there are a number of loose p...

Страница 13: ...ic cap assembly by pushing it back into the top of the container Tighten the wing nut to seal the container 4 Connect the oxygen supply to the outlet fitted in the cap Assembling the Flexible Hoses to...

Страница 14: ...iece is a tight fit into the hose and this is as it should be Hold the hose and tilt the T piece back slightly Push 3 4 of the T piece inside the hose gently wiggle and rotate both parts to fully inse...

Страница 15: ...old together whilst breathing Option 2 BX0978 Disposable Breathing Filter fit the accessory adaptor MXM0894 into the T piece Push the disposable filter into the adaptor Fitting the Pen Arm for Recordi...

Страница 16: ...SPIROMETER SM 1 CLIFTON RANGE 1 YEAR WARRANTY...

Страница 17: ...SPIROMETER SM 1 CLIFTON RANGE 1 YEAR WARRANTY...

Страница 18: ...SPIROMETER SM 1 CLIFTON RANGE 1 YEAR WARRANTY...

Страница 19: ...SPIROMETER SM 1 CLIFTON RANGE 1 YEAR WARRANTY...

Страница 20: ...SPIROMETER SM 1 CLIFTON RANGE 1 YEAR WARRANTY...

Страница 21: ...BX0854 Issue 3 April 2020...

Отзывы: