SPIROMETER SM-1
CLIFTON RANGE® - 1 YEAR WARRANTY 

UNPACKING

Remove the product from its packaging and retain over the warranty period.  Contents 

consist of:

• 

Spirometer Body

• 

Float Assembly

• 

Pen Arm Assembly

• 

Nylon Net Bag

• 

Corrugated Tube (x2)

• 

Snapper Clip (x2)

• 

Mouth Piece

• 

Nose Clip

• 

Instruction Manual

SETTING UP (SEE APPENDIX A FOR FURTHER INFORMATION)

1.  Place the float in position, ensuring that the pivot pins are correctly located in the 

dimples( top of the tank).

2.  Screw the counter balance weight fully into the back of the float support but do not 

adjust the float until the chamber has been filled with water.

3.  Connect the flexible hoses to the ports of the Spirometer. When new these may be a 

tight fit (lubricate with water). Fit the clips to secure. Note the way the valves in the 

T piece operates and connect the corugated hoses to the T piece so that expired air 

enters the spirometer via the port labelled IN and expired air from the float leaves 

via the port labelled OUT. DO NOT use oil or grease as a lubricant anywhere in the 

system as oxygen can form explosive compounds with these materials.

4.  Immerse the mouth piece into sterilizing solution. Spare mouth pieces are available 

from your distributor (BX0612 Disposable Mouth Piece).

5.  If the results are to be recorded on a kymograph, fit the pen arm and clamp a fine 

fibre tipped pen into the holder. Arrange the kymograph so that the pen writes 

properly over the full height of the chart. If the full spirometer capacity of 7 litres is 

to be used, a chart and 200mm is required. When fitted 10mm of vertical movement 

of the pen corresponds to approx 500ml change in volume in the spirometer.

6.  Fill the tank with water almost to the very top.

7.  Calibrate the recording device to known volume of air for the “Suggested 

Investigations”. 

8.  Ensure the float is in the lowest position, set the valve to ATMOSPHERE. The 

kymograph should produce a continuous horizontal lineand dataloggers will record 

zero motion. This corresponds to zero volume.

9.  To introduce a measured volume of air, fill a flask with an accurately measured 

volume of water (500ml). Mark the level of water on the neck of the flask. Empty 

the water, fit a two hole bung with one tube connected to the oxygen inlet of the 

Содержание Clifton Range SM-1

Страница 1: ...High quality products at affordable prices SPIROMETER SM 1...

Страница 2: ......

Страница 3: ...should only be used under the direct supervision of a teacher who is experienced in their correct use Teachers are reminded that they have a special responsbility when pupils are used as the subject o...

Страница 4: ...designed to offer minimum resistance to air flow This means greater comfort for the user and more realistic results The spirometer enables rate of breathing volumes of breaths and consumption of oxy...

Страница 5: ...anywhere in the system as oxygen can form explosive compounds with these materials 4 Immerse the mouth piece into sterilizing solution Spare mouth pieces are available from your distributor BX0612 Dis...

Страница 6: ...res turn off the supply set the valve to SPIROMETER and press down on the float This forces the oxygen out through the mouthpiece and flushes the air out of the system Do not allow oxygen to be discha...

Страница 7: ...by the amplitude of the trace 4 Respiratory rate the number of breathes per minute or more formally the number of movements indicative of inspiration and expiration is usually determined by counting...

Страница 8: ...new condition Descale by adding 1 litre of vinegar to water and leave for an hour Empty and brush the lime away Rinse thoroughly afterwards The stainless steel crevice free tank should provide years o...

Страница 9: ...y materials or workmanship in any goods manufactured supplied by Nickel Electro Ltd which under proper and normal conditions of use may develop within a period of one years from the date of delivery 3...

Страница 10: ...for specialist engineers currently in force 7 In the event of replacement with a new or reconditioned model the replacement unit will continue the warranty period of the original equipment 8 If any g...

Страница 11: ...rbalance weight is supplied Slide the weight onto the rod and tighten the thumb screw This is the counterweight assembly for the float 2 The large white moulding is the float assembly a metal frame is...

Страница 12: ...rew is fitted Each one locates into a dimple on the top rim of the tank The float assembly will see saw adjust the counter balance weight accordingly 5 In a polythene bag there are a number of loose p...

Страница 13: ...ic cap assembly by pushing it back into the top of the container Tighten the wing nut to seal the container 4 Connect the oxygen supply to the outlet fitted in the cap Assembling the Flexible Hoses to...

Страница 14: ...iece is a tight fit into the hose and this is as it should be Hold the hose and tilt the T piece back slightly Push 3 4 of the T piece inside the hose gently wiggle and rotate both parts to fully inse...

Страница 15: ...old together whilst breathing Option 2 BX0978 Disposable Breathing Filter fit the accessory adaptor MXM0894 into the T piece Push the disposable filter into the adaptor Fitting the Pen Arm for Recordi...

Страница 16: ...SPIROMETER SM 1 CLIFTON RANGE 1 YEAR WARRANTY...

Страница 17: ...SPIROMETER SM 1 CLIFTON RANGE 1 YEAR WARRANTY...

Страница 18: ...SPIROMETER SM 1 CLIFTON RANGE 1 YEAR WARRANTY...

Страница 19: ...SPIROMETER SM 1 CLIFTON RANGE 1 YEAR WARRANTY...

Страница 20: ...SPIROMETER SM 1 CLIFTON RANGE 1 YEAR WARRANTY...

Страница 21: ...BX0854 Issue 3 April 2020...

Отзывы: