2. Firmware update
Reducing Vaporization Mercury Analyzer Mercury RA-4500
3
NIC-704-2005-01
2.
Firmware update
- Before updating the firmware, make sure that it is compatible with the version of
the personal computer software.
2.1.
Procedure for updating firmware
Since RA4Win and the firmware have a rewriting function for the firmware, it can be updated
easily.
Click “System” and “Firmware Update” of RA-4win.
The screen shown above appears. This screen will be used to update the firmware.
Follow the procedure described below:
(1) Make a firmware file read into the personal computer.
(2) Write the read data to the flash memory on the mainboard of the RA4.
(3) Turn off and on the main unit to enable the update.
Warning