7.5. Measurement of Unknown samples 

(1)  Enter  the  sample  name  and  the  sample  amount.  Select  the  sample  form  and  unit  etc  on  the 

“Table Condition” by clicking          . 

(2) Select the heating method in “Method Number.” If the moisture content is known as for solid 

samples, enter it in “MOIST [%]” so that the concentration is calculated as “dry weight.” 

(3) After checking 

“Sample Number” with

, click            or press “F5” to start measuring. 

 

 

 

 

 

 

 

Selection for (Heating) Method

   

Select a heating method

 

(12 Methods)

 

can be registered in “Run - Method set” as Heating Method.   

5 Methods are installed by default as described below. 

Heating Method table 

Method 

No, 

Kind of sample 

Heating Method 

Typical 

amount 

Liquid sample /Standard solution 
Liquid sample with additive B 

STD 

~

200μL 

Biological sample

 

(Solid) 

Fish tissue, plant tissue 

ORGANISM 

(SOLID) 

~200 mg 

Soil, sediment & sludge 

SOIL & SLUDGE 

~200 mg 

Urine, Blood 

URINE & BLOOD 

~

200μL 

Plastics 

PLASTICS 

~100 mg 

Minerals 
Coal, coal ash, metal and mineral 

MINERAL 

~200 mg 

Waste water   

WASTE WATER 

~2

00μL 

Purge 

PURGE 

 

Do  not  measure  the  sample  digested  by  HNO

3

  directly.  This  sample  may  generate  heat  on  the 

activated carbon filter. Dilute it less than 5wt% of HNO

3

.

 

※ 

As for HCl solution, dilute it less than 1wt%.

 

 

The above “Heating Methods” may not cover the samples which contain a lot of halogens or burn 

quickly. 

      When the check standard after measuring is lower, reduce the sample size. 

  Dilute  or  reduce  the  sample  size  if  it  contains  high  mercury.  Otherwise  it  contaminates  the 

plumbing. 

 As for the liquid samples c

ontaining organic mercury, use “Additive B” on the boat. 

Содержание MERCURY/MA-3000

Страница 1: ...MERCURY MA 3000 Simplified Operation manual NIC 600 2233 06 Nippon Instruments Corporation...

Страница 2: ...hat incorrect handling could result in death or severe injury to the operator Caution Indicates that incorrect handling could result in moderate or minor injury to the operator or damage to property 1...

Страница 3: ...the discharge If you don t turn on MA 3000 for a long time the sample heating tube might be obstructed or break by moisture absorption If you don t turn on MA 3000 more than 2 weeks please turn on th...

Страница 4: ...essage will be displayed if abnormal condition is observed Software will be terminated when OK is clicked Turn off MA 3000 and then fixing them 4 After the check is completed heat conditioning start H...

Страница 5: ...method for STD The default method is already set for standard solution but selects the proper method for using SRM Standard Reference Materials Select Flow rate 0 1L min High sensitivity 1pg 10ng 0 2L...

Страница 6: ...Solid or Liquid Select a sample feature This is not related to Measuring method Calibration curve Usually please select y ax blank 7 Measurements 7 1 Registration of standard Click the tab of STD 1 C...

Страница 7: ...Click the icon and switch to the Analyze screen Review the calibration curve in CALIB tab When some abnormalities are observed on a calibration curve all values of the calibration become red and WARNI...

Страница 8: ...e B or L cysteine solution not to vaporize during standby in the sample changer Add small amount of additive B in the boat bottom or add 10 L of 100ppm L cysteine solution When you use additive B plea...

Страница 9: ...Standard solution Liquid sample with additive B STD 200 L 1 Biological sample Solid Fish tissue plant tissue ORGANISM SOLID 200 mg 2 Soil sediment sludge SOIL SLUDGE 200 mg 3 Urine Blood URINE BLOOD...

Страница 10: ...t of sample or to change the heating conditions Inject 200 L of Distilled Water for several times if high concentration samples are measured Inject 200 L of Standard Solution repeatedly if the sensiti...

Страница 11: ...turn on MA 3000 for a long time the sample heating tube might be obstructed by moisture absorption If you don t turn on MA 3000 more than 2 weeks please turn on the MA 3000 power switch and do conditi...

Страница 12: ...33 06 11 Nippon Instruments Corporation 14 8 Akaoji cho Takatsuki shi Osaka 569 1146 Japan Please contact the distributor or the following Technical Service E mail info nic rigaku co jp URL https www...

Отзывы: