Pretreatment of sample boat 

Sample  boats  absorb  mercury  in  the  air  and  a  blank  becomes  high.  In  order  to  measure  with 

sufficient accuracy, heat-treat to remove mercury sample boats before measurement.   

The way of heat-treatment 

(a)  Sample boats are repeatedly and measured on MA-3000 to confirm low enough.   

(b)  Sample boats are put into a ceramic crucible, which is heated at 750 degrees for 1 hour or more. 

*At first use, 

don’t wash the sample boats before heat treatment. Performance may be impaired. 

 

5. Setting for Measurement Condition 

(

Click            and “Measurement Condition” is displayed. 

Select  the  proper  method  for  “STD.”  The  default 

method  is  already  set  for  standard  solution,  but 
selects the proper method for using SRM (Standard 
Reference Materials). 

 

Select 

“Flow rate”.

 

 

0.1L /min: High sensitivity   

(1pg~10ng) 

0.2L/min: usual                       

 

0.5L/min: high concentration   

(Linear range up to 10,000ng) 

 

Enter the proper values in each item of “STD Check.” 

 

 

In  thermal  decomposition  mercury  analyzing,  approx 
1%  of  the  prior  measured  value  is  detected  as  a 
carryover.  If  a  low-concentration  sample  is  measured 
after  a  high  concentration  sample,  it  will  be  affected 
with previous carryover. 
 
“Blank Check” function provides to automatically lower 
a blank. 

 

 

 

 

 

 

 

 

Please  refer  this  table  which  number  should  be 

entered.

   

Typical value (0.2L/min) 

ABS(LOW) 

ABS(HIGH) 

0.1ng 

0.002 

10ng 

0.001 

1ng 

0.02 

100ng 

0.01 

5ng 

0.1 

1,000ng 

0.1 

10ng 

0.2 

5,000ng 

0.4 

20ng 

0.3 

10,000ng 

0.6 

Содержание MERCURY/MA-3000

Страница 1: ...MERCURY MA 3000 Simplified Operation manual NIC 600 2233 06 Nippon Instruments Corporation...

Страница 2: ...hat incorrect handling could result in death or severe injury to the operator Caution Indicates that incorrect handling could result in moderate or minor injury to the operator or damage to property 1...

Страница 3: ...the discharge If you don t turn on MA 3000 for a long time the sample heating tube might be obstructed or break by moisture absorption If you don t turn on MA 3000 more than 2 weeks please turn on th...

Страница 4: ...essage will be displayed if abnormal condition is observed Software will be terminated when OK is clicked Turn off MA 3000 and then fixing them 4 After the check is completed heat conditioning start H...

Страница 5: ...method for STD The default method is already set for standard solution but selects the proper method for using SRM Standard Reference Materials Select Flow rate 0 1L min High sensitivity 1pg 10ng 0 2L...

Страница 6: ...Solid or Liquid Select a sample feature This is not related to Measuring method Calibration curve Usually please select y ax blank 7 Measurements 7 1 Registration of standard Click the tab of STD 1 C...

Страница 7: ...Click the icon and switch to the Analyze screen Review the calibration curve in CALIB tab When some abnormalities are observed on a calibration curve all values of the calibration become red and WARNI...

Страница 8: ...e B or L cysteine solution not to vaporize during standby in the sample changer Add small amount of additive B in the boat bottom or add 10 L of 100ppm L cysteine solution When you use additive B plea...

Страница 9: ...Standard solution Liquid sample with additive B STD 200 L 1 Biological sample Solid Fish tissue plant tissue ORGANISM SOLID 200 mg 2 Soil sediment sludge SOIL SLUDGE 200 mg 3 Urine Blood URINE BLOOD...

Страница 10: ...t of sample or to change the heating conditions Inject 200 L of Distilled Water for several times if high concentration samples are measured Inject 200 L of Standard Solution repeatedly if the sensiti...

Страница 11: ...turn on MA 3000 for a long time the sample heating tube might be obstructed by moisture absorption If you don t turn on MA 3000 more than 2 weeks please turn on the MA 3000 power switch and do conditi...

Страница 12: ...33 06 11 Nippon Instruments Corporation 14 8 Akaoji cho Takatsuki shi Osaka 569 1146 Japan Please contact the distributor or the following Technical Service E mail info nic rigaku co jp URL https www...

Отзывы: