The row width of printing is being interlocked with a setup of Table screen. If the row width at the 

time of printing is narrow, please adjust on the screen. And, the row will not be printed if width of 

a row is made into zero. 

 

10. Termination 

(

)  Press the upper right button        of MA3win. 

(

)  Close the Windows and turn off the power of PC and Printer. 

(

)  Power off MA-3000. 

(4)

  Discharge liquid of a drain tank.

 

(

)  Close the main cock of the Oxygen gas cylinder. 

 

 

 

*

If  you  don’t  turn-on  MA-3000  for  a  long  time,  the  sample  heating  tube  might  be  obstructed  by 

moisture absorption. 

If you don’t turn-on MA-3000 more than 2 weeks, please turn on the MA-3000 power switch and do 

conditioning. (Conditioning process start automatically after checking the flow path) 

Otherwise, please dismount the sample heating tube and store it in a desiccator to prevent moisture 

absorption. 

 

 

11.   Errors and warning 

Measurement  can  continue  even  if  a  warning  message  has  been  displayed.  However,  if  an  error 

message is displayed, measurement will be disabled until the cause of the error has been solved. 

Please check messages and contact our service. 

 

Low     

HIGH 

Number 
of profile 

Содержание MERCURY/MA-3000

Страница 1: ...MERCURY MA 3000 Simplified Operation manual NIC 600 2233 06 Nippon Instruments Corporation...

Страница 2: ...hat incorrect handling could result in death or severe injury to the operator Caution Indicates that incorrect handling could result in moderate or minor injury to the operator or damage to property 1...

Страница 3: ...the discharge If you don t turn on MA 3000 for a long time the sample heating tube might be obstructed or break by moisture absorption If you don t turn on MA 3000 more than 2 weeks please turn on th...

Страница 4: ...essage will be displayed if abnormal condition is observed Software will be terminated when OK is clicked Turn off MA 3000 and then fixing them 4 After the check is completed heat conditioning start H...

Страница 5: ...method for STD The default method is already set for standard solution but selects the proper method for using SRM Standard Reference Materials Select Flow rate 0 1L min High sensitivity 1pg 10ng 0 2L...

Страница 6: ...Solid or Liquid Select a sample feature This is not related to Measuring method Calibration curve Usually please select y ax blank 7 Measurements 7 1 Registration of standard Click the tab of STD 1 C...

Страница 7: ...Click the icon and switch to the Analyze screen Review the calibration curve in CALIB tab When some abnormalities are observed on a calibration curve all values of the calibration become red and WARNI...

Страница 8: ...e B or L cysteine solution not to vaporize during standby in the sample changer Add small amount of additive B in the boat bottom or add 10 L of 100ppm L cysteine solution When you use additive B plea...

Страница 9: ...Standard solution Liquid sample with additive B STD 200 L 1 Biological sample Solid Fish tissue plant tissue ORGANISM SOLID 200 mg 2 Soil sediment sludge SOIL SLUDGE 200 mg 3 Urine Blood URINE BLOOD...

Страница 10: ...t of sample or to change the heating conditions Inject 200 L of Distilled Water for several times if high concentration samples are measured Inject 200 L of Standard Solution repeatedly if the sensiti...

Страница 11: ...turn on MA 3000 for a long time the sample heating tube might be obstructed by moisture absorption If you don t turn on MA 3000 more than 2 weeks please turn on the MA 3000 power switch and do conditi...

Страница 12: ...33 06 11 Nippon Instruments Corporation 14 8 Akaoji cho Takatsuki shi Osaka 569 1146 Japan Please contact the distributor or the following Technical Service E mail info nic rigaku co jp URL https www...

Отзывы: