Nic MERCURY/MA-3000 Скачать руководство пользователя страница 1

 

 

 

 

MERCURY/MA-3000 

 

Simplified

 Operation manual 

 

 

 

 

NIC-600-2233-06 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Nippon Instruments Corporation

 

Содержание MERCURY/MA-3000

Страница 1: ...MERCURY MA 3000 Simplified Operation manual NIC 600 2233 06 Nippon Instruments Corporation...

Страница 2: ...hat incorrect handling could result in death or severe injury to the operator Caution Indicates that incorrect handling could result in moderate or minor injury to the operator or damage to property 1...

Страница 3: ...the discharge If you don t turn on MA 3000 for a long time the sample heating tube might be obstructed or break by moisture absorption If you don t turn on MA 3000 more than 2 weeks please turn on th...

Страница 4: ...essage will be displayed if abnormal condition is observed Software will be terminated when OK is clicked Turn off MA 3000 and then fixing them 4 After the check is completed heat conditioning start H...

Страница 5: ...method for STD The default method is already set for standard solution but selects the proper method for using SRM Standard Reference Materials Select Flow rate 0 1L min High sensitivity 1pg 10ng 0 2L...

Страница 6: ...Solid or Liquid Select a sample feature This is not related to Measuring method Calibration curve Usually please select y ax blank 7 Measurements 7 1 Registration of standard Click the tab of STD 1 C...

Страница 7: ...Click the icon and switch to the Analyze screen Review the calibration curve in CALIB tab When some abnormalities are observed on a calibration curve all values of the calibration become red and WARNI...

Страница 8: ...e B or L cysteine solution not to vaporize during standby in the sample changer Add small amount of additive B in the boat bottom or add 10 L of 100ppm L cysteine solution When you use additive B plea...

Страница 9: ...Standard solution Liquid sample with additive B STD 200 L 1 Biological sample Solid Fish tissue plant tissue ORGANISM SOLID 200 mg 2 Soil sediment sludge SOIL SLUDGE 200 mg 3 Urine Blood URINE BLOOD...

Страница 10: ...t of sample or to change the heating conditions Inject 200 L of Distilled Water for several times if high concentration samples are measured Inject 200 L of Standard Solution repeatedly if the sensiti...

Страница 11: ...turn on MA 3000 for a long time the sample heating tube might be obstructed by moisture absorption If you don t turn on MA 3000 more than 2 weeks please turn on the MA 3000 power switch and do conditi...

Страница 12: ...33 06 11 Nippon Instruments Corporation 14 8 Akaoji cho Takatsuki shi Osaka 569 1146 Japan Please contact the distributor or the following Technical Service E mail info nic rigaku co jp URL https www...

Отзывы: