Battery operating temperature range
Energizer E91.................................................–18 °C to 55 °C
Duracell MN1500 ..........................................–20 °C to 54 °C
Energizer L91.................................................–40 °C to 60 °C
Voltage range .........................................................3.6 to 7.5 V
Power consumption
1
Sleep mode.............................................................0.3 mW
Note
Power consumption and battery life varies per application. Refer to the
section for examples that show how battery performance is impacted by settings like samples per
waveform and sample interval.
External Power
Caution
You
must
use a UL Listed ITE power supply marked LPS with the NI WSN-3214. The
power supply must also meet any safety and compliance requirements for the country of use.
Voltage range .........................................................5 to 30 V
Power input mating connector ...............................2-position mini-combicon,
Phoenix Contact part number: 1714977
Power consumption
2
Sleep mode.....................................................1 mW
60 second sample interval..............................4 mW at 12 V
5 second sample interval ................................40 mW at 12 V
Router mode
3
.................................................200 mW at 12 V
Physical Characteristics
Screw-terminal wiring ...........................................14 to 24 AWG wire
Torque for screw terminals ....................................0.2 to 0.25 N
·
m
Dimensions ............................................................Refer to Figure 2 for device dimensions
Weight ....................................................................Approx. 242 g (8.5 oz)
Weight with antenna .............................................Approx. 256 g (9 oz)
Calibration
You can obtain the calibration certificate and information about calibration services for the
NI WSN-3214 at
ni.com/calibration
.
Calibration interval ................................................3 years
Safety Standards
Caution
Do not operate the NI WSN-3214 in a manner not specified in this user guide. Product
misuse can result in a hazard. You can compromise the safety protection built into the product if the
product is damaged in any way. If the product is damaged, return it to National Instruments for repair.
1
Device executing NI-WSN firmware.
2
Device executing NI-WSN firmware. 4 channels, 10 S/waveform, 50/60 Hz high rejection.
3
Router connected directly to a NI WSN gateway. 1 second sample interval.