![NI USB-9263 Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/ni/usb-9263/usb-9263_user-manual_1685553011.webp)
©
National Instruments Corporation
11
NI USB-9263 User Guide and Specifications
Physical Characteristics
Screw terminal wiring ............................ 12 to 24 AWG copper conductor
wire with 10 mm (0.39 in.) of
insulation stripped from the end
Torque for screw terminals .................... 0.5 to 0.6 N · m (4.4 to 5.3 lb · in.)
Ferrules................................................... 0.25 mm
2
to 2.5 mm
2
Weight .................................................... Approx. 302 g (10.6 oz)
Safety
If you need to clean the module, wipe it with a dry towel.
Safety Voltages
Channel-to-COM ................................... ±11 V max
Isolation
Channel-to-channel ......................... None
Channel-to-earth ground
Continuous............................... 250 V
rms
,
Measurement Category II
Withstand................................. 2,300 V
rms
, verified by a 5 s
dielectric withstand test
Measurement Category II is for measurements performed on circuits
directly connected to the electrical distribution system. This category refers
to local-level electrical distribution, such as that provided by a standard
wall outlet (for example, 115 V for U.S. or 230 V for Europe). Do
not
connect to signals or use for measurements within Measurement
Categories III or IV.
Safety Standards
This product is designed to meet the requirements of the following
standards of safety for electrical equipment for measurement, control,
and laboratory use:
•
IEC 61010-1, EN 61010-1
•
UL 61010-1, CSA 61010-1
Note
For UL and other safety certifications, refer to the product label or the