Chapter 1 Installing and Configuring the SCXI-1127
3. Connect the SH9MD-9MD cable from the AUX OUT connector of the
first chassis to the AUX IN connector of the next chassis.
4. Repeat steps 1 through 3 for each additional chassis.
Note A n SC XI-1127 is required in slot 4 to establish communications with the chassis.
I f slot 4 is empty, the system will not operate. It is this module that you must specify
in the configuration utility as the cabled module to the DMM.
I f a chassis is configured with less than four slots of the HVAB-backplane adapter,
use the right-most available HVAB connector to extend to the next chassis. In this
instance, right-most means the slot closest to slot 1 when the chassis is viewed from
the back.
4-Slot and 12-Slot Multi-Chassis Configuration
In addition to the items needed for a single chassis system, you need the
following items to install this configuration:
• S C X I - 1 3 5 8 multi-chassis expansion kit
HV8-HV8 cable
— S H 9 M D - 9 M D cable
- T h r e e 8-position HVAB plugs
-
O n e
8
-slot
HVAB-backplane adapter
- O n e 2-slot HVAB-backplane adapter
- T w o 1-slot HVAB-backplane adapters