NI PXIe-5673E Скачать руководство пользователя страница 2

Programming the NI 5673E.................................................................................................... 19

NI-RFSG Examples........................................................................................................ 19
Generating a Signal Using the NI-RFSG Soft Front Panel.............................................20
Building a Basic NI-RFSG Application..........................................................................20

Troubleshooting...................................................................................................................... 24

Why Is the ACCESS LED Off When the Chassis is On?...............................................24
What Should I Do if the NI 5673E Doesn't Appear in MAX?....................................... 24
What Should I Do if the Module Fails the Self-Test?.....................................................25
What Should I Do if the Thermal Shutdown Error Appears?.........................................25

Where to Go Next................................................................................................................... 26
Worldwide Support and Services............................................................................................ 26

Electromagnetic Compatibility Guidelines

This product was tested and complies with the regulatory requirements and limits for
electromagnetic compatibility (EMC) stated in the product specifications. These requirements
and limits are designed to provide reasonable protection against harmful interference when the
product is operated in the intended operational electromagnetic environment.

This product is intended for use in industrial locations. However, harmful interference may
occur in some installations, when the product is connected to a peripheral device or test object,
or if the product is used in residential or commercial areas. To minimize interference with
radio and television reception and prevent unacceptable performance degradation, install and
use this product in strict accordance with the instructions in the product documentation.

Furthermore, any modifications to the product not expressly approved by National Instruments
could void your authority to operate it under your local regulatory rules.

Caution

 

To ensure the specified EMC performance, operate this product only with

shielded cables and accessories.

Caution

 

To ensure the specified EMC performance, the length of all I/O cables

must be no longer than 3 m (10 ft) with the exception of cables connected to the
RF OUT and PLS MOD ports, which must be no longer than 10 m (30 ft).

Verifying the System Requirements

To use the NI-RFSG instrument driver, your system must meet certain requirements.

For more information about minimum system requirements, recommended system, and
supported application development environments (ADEs), refer to the product readme, which
is available on the driver software media or online at 

ni.com/updates

.

Unpacking the Kit

Caution

 

To prevent electrostatic discharge (ESD) from damaging the device,

ground yourself using a grounding strap or by holding a grounded object, such as
your computer chassis.

2

  | 

ni.com

  | 

NI PXIe-5673E Getting Started Guide

Содержание PXIe-5673E

Страница 1: ...ss NI 5673E documentation navigate to Start All Programs National Instruments NI RFSG Documentation The specifications document for your device is installed with the driver software Refer to ni com ma...

Страница 2: ...ze interference with radio and television reception and prevent unacceptable performance degradation install and use this product in strict accordance with the instructions in the product documentatio...

Страница 3: ...or any other sign of damage Caution Never touch the exposed pins of connectors Notify NI if the device appears damaged in any way Do not install a damaged device 3 Unpack any other items and document...

Страница 4: ...1 AWG Module 3 NI PXIe 5650 5651 5652 LO Source Module 4 Four Short Rigid SMA to SMA Cables Labeled A Part Number 195947 01 5 Medium Semi Rigid SMA to SMA Cable Labeled B Part Number 198774 01 6 Semi...

Страница 5: ...73E in meets the following specifications Operating ambient temperature IEC 60068 2 1 IEC 60068 2 2 0 C to 55 C Operating relative humidity IEC 60068 2 56 10 to 90 noncondensing Maximum altitude 2 000...

Страница 6: ...module in the slot immediately to the right of the NI 5450 5451 AWG module and you must install the NI 5650 5651 5652 LO source module in the slot immediately to the right of the NI 5611 I Q modulator...

Страница 7: ...rdware Module 3 Ejector Handle in Down Unlatched Position 9 Latch the module in place by pulling up on the ejector handle 10 Secure the device front panel to the chassis using the front panel mounting...

Страница 8: ...id SMA to SMA Cable Labeled A 3 Semi Rigid SMA to SMA Cable Labeled B 4 50 Termination Load 1 Use the Cable A center pin to connect the CH 1 Q connector of the NI 5450 5451 front panel to the Q connec...

Страница 9: ...e or effort If a cable bends due to incorrect alignment remove the cable from the front panel connector and reconnect the cable center pin on the SMA cable to the SMA connector 9 Carefully complete ti...

Страница 10: ...in a static free environment If you are using an active device such as a preamplifier or switch routed to the NI 5673E ensure no signal transients are sourced to the NI 5673E NI 5611 I Q Modulator Mod...

Страница 11: ...output used for combining multiple devices with one common LO source NI 5650 5651 5652 Attach a 50 termination load to LO OUT when the connector is not in use I Connects to the CH 0 I connector on the...

Страница 12: ...icates one of the following error states Reverse power protection fault The RF OUT connector is disabled to protect the device from the power being driven back into the RF OUT connector I Q overvoltag...

Страница 13: ...l Table 4 Device Front Panel Icon Definition Refer to the user documentation for required maintenance measures to ensure user safety and or preserve the specified EMC performance NI PXIe 5673E Getting...

Страница 14: ...and trigger sources This connector can connect to the CLK OUT connector on an additional NI 5450 5451 front panel PFI 1 Accepts a trigger from an external source that can start or step through wavefo...

Страница 15: ...OFF The module is not armed or triggered AMBER The module is armed and waiting for a Start Trigger GREEN The module has received a Start Trigger and is generating a waveform RED The module has detecte...

Страница 16: ...CCESS ACTIVE Table 7 Device Front Panel Icon Definitions Refer to the user documentation for required maintenance measures to ensure user safety and or preserve the specified EMC performance If this d...

Страница 17: ...Express power rail AMBER The module is being accessed Accessed means that the device setup registers are being written to in order to control the device GREEN The module is ready to be programmed by N...

Страница 18: ...re modules that comprise your device a Select the NI 5611 that is identified as not configured in the configuration tree b In the Associated Devices section select the appropriate module from each sys...

Страница 19: ...he NI 5673E in MAX NI RFSG operates the modules as a single instrument LabWindows CVI Available at Program Files IVI Foundation IVI Drivers niRFSG Microsoft Visual C C Add all required include and lib...

Страница 20: ...ubleshooting section of this document if an ACTIVE LED does not turn on or if the NI RFSG SFP generates an error During signal generation the ACTIVE LED on each of the NI 5673E hardware modules illumi...

Страница 21: ...n the VI front panel power level dBm control enter 0 In the frequency Hz control enter 100M 100 MHz 12 In the VI front panel resource name control enter the NI 5611 I Q modulator device name that you...

Страница 22: ...utput of the status element to an input of the Or function 9 Right click the unused input of the Or function and select Create Control to create a Boolean control 10 In the VI front panel right click...

Страница 23: ...odule name specified in MAX in the resource name control 4 Click the Run button on the toolbar to initiate sine wave generation 5 Click the VI front panel STOP button to stop sine wave generation You...

Страница 24: ...not reinstall a damaged device 4 Reinstall the module in a different chassis slot 5 Power on the chassis 6 Verify that the device appears in MAX 7 Reset the device in MAX and perform a self test What...

Страница 25: ...in a different slot 5 Power on the chassis 6 Perform the self test again What Should I Do if the Thermal Shutdown Error Appears The thermal shutdown error appears when device temperatures exceed safe...

Страница 26: ...software EXPLORE LEARN CREATE DISCOVER RF Solutions ni com rf Services ni com services NI Community ni com community Located at ni com gettingstarted Located at ni com manuals Located using the NI Exa...