![NI NI-9853 Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/ni/ni-9853/ni-9853_manual_1685514012.webp)
Withstand
500 V
rms
, verified by a 5 s dielectric withstand test
Port-to-earth ground
Continuous
60 VDC, Measurement Category I up to 5,000 m altitude
Withstand
500 V
rms
, verified by a 5 s dielectric withstand test
Measurement Category I is for measurements performed on circuits not directly
connected to the electrical distribution system referred to as MAINS voltage. MAINS
is a hazardous live electrical supply system that powers equipment. This category is
for measurements of voltages from specially protected secondary circuits. Such
voltage measurements include signal levels, special equipment, limited-energy
parts of equipment, circuits powered by regulated low-voltage sources, and
electronics.
Note
Do not connect the NI-9853 to signals or use for measurements
within Measurement Categories II, III, or IV.
The maximum voltage that can be applied or output between any port or V
SUP
terminal and a COM terminal without creating a safety hazard.
Hazardous Locations
U.S. (UL)
Class I, Division 2, Groups A, B, C, D, T4; Class I, Zone 2, AEx nA IIC T4
Canada (C-UL)
Class I, Division 2, Groups A, B, C, D, T4; Class I, Zone 2, Ex nA IIC T4
Europe (DEMKO)
Ex nA IIC T4 Gc
Safety and Hazardous Locations Standards
This product is designed to meet the requirements of the following electrical
equipment safety standards for measurement, control, and laboratory use:
■
IEC 61010-1, EN 61010-1
■
UL 61010-1, CSA 61010-1