Calibrating FieldPoint AI Modules with Calibration Executive
12
ni.com
Connecting the Calibrator to an FP-RTD-122 Module
If you are calibrating the FP-RTD-122 module, make the following
connections as shown in Figure 8.
1.
Connect the positive output (ohms) and positive sense (ohms) of the
calibrator to terminal 1 of the calibration base.
2.
Connect the negative output (ohms) and the negative sense (ohms) of
the calibrator to terminal 18 of the calibration base.
3.
On the calibration terminal base, connect terminal 2 to terminal 18.
While the Calibration Executive procedure runs, you are prompted to make
additional connections between the calibrator and the module you are
calibrating. Follow the onscreen instructions at each prompt.
Figure 8.
FP-RTD-122 Connections to Calibrator
1
FP-1000 Network Module
2
TB-Cal Calibration Terminal Base
3
Connect terminal 2 to terminal 18
4
Connect to positive output of calibrator
5
Connect to positive sense of calibrator
6
Connect to negative output of calibrator
7
Connect to negative sense of calibrator
8