Calibrating FieldPoint AI Modules with Calibration Executive
Connecting the Calibrator to an FP-AI-1
xx
Module
If you are calibrating an FP-AI-100, FP-AI-110, or FP-AI-111 module,
make the connections shown in Figure 7.
1.
Connect the positive output (current) of the calibrator to terminal 2 of
the calibration terminal base.
2.
Connect the negative output (current) of the calibrator to terminal 18 of
the calibration terminal base.
Figure 7.
FP-AI-1
xx
Connections to Calibrator for Current Offset Testing
1
FP-1000 Network Module
2
FP-TB-CAL Calibration Terminal Base
3
Connect to positive output (current) of calibrator
4
Connect to negative output (current) of calibrator
5
Connect to DC V power supply
6
Connect RS-232 serial cable to COM port
7
Baud Rate Switches