Calibrating FieldPoint AI Modules with Calibration Executive
10
ni.com
Caution
You
must
close FieldPoint Explorer and keep it closed during the remainder of
this procedure to prevent a system crash.
Connecting the Calibrator to an FP-TC-120 Module
If you are calibrating the FP-TC-120, make the following connections as
shown in Figure 6:
1.
Connect the positive output (voltage) of the calibrator to terminal 1 of
the calibration terminal base
2.
Connect the negative output (voltage) of the calibrator to terminal 2 of
the calibration terminal base.
Figure 6.
FP-TC-120 Connections to Calibrator