on Proced
ure
SMA (m)-to-SMA (m)
cable
—
All procedures
Frequency range: DC to 2 GHz
Impedance: 50
Ω
Maximum length: 1 meter
SMA (f)-to-N (m)
adapter
Fairview Microwave
SM4226
All procedures
Frequency range: DC to 2 GHz
Impedance: 50
Ω
Return loss:
≥
23 dB
SMA (m)-to-N (f)
adapter
Huber+Suhner
33_SMA_N-50-1/1-_UE or
Fairview Microwave
SM4241
Test system characterization
Verifying absolute amplitude accuracy
Adjusting absolute amplitude accuracy
Frequency range: DC to 2 GHz
Impedance: 50
Ω
Return loss:
≥
23 dB
SMA (f)-to-N (f)
adapter
Huber+Suhner
31_3-SMA-50-1/1-_UE or
Fairview Microwave
SM4236
Test system characterization
Frequency range: DC to 2 GHz
Impedance: 50
Ω
Return loss:
≥
23 dB
3.5 mm
(m)-to-3.5 mm (m)
adapter
Huber+Suhner
32_PC35-50-0-2/199_NE or
Fairview Microwave
SM4960
Test system characterization
Verifying absolute amplitude accuracy
Adjusting absolute amplitude accuracy
Frequency range: DC to 2 GHz
Impedance: 50
Ω
Return loss:
≥
30 dB
3.5 mm (f)-to-
3.5 mm (f) adapter
Huber+Suhner
31_PC35-50-0-1/199_UE
Test system characterization
Frequency range: DC to 2 GHz
Impedance: 50
Ω
Return loss:
≥
28 dB
Table 1.
NI 5624R Required Equipment Specifications for NI 5624R Calibration (Continued)
Equipment
Recommended Model
Measured Specification
Minimum Requirements