44
Troubleshooting
Troubleshooting
To help you understand some of the problems that might occur, here is a list of potential faults
with some solutions. For assistance in setting up, using or maintaining the equipment please
contact Odstock Medical Limited or their local representative.
1. No stimulation response to the footswitch but response to the test button
Fault
Action
The stimulator may be paused
Press the pause button
Faulty footswitch
Replace footswitch
Faulty footswitch lead
Replace footswitch lead
Lead not fully inserted/connected
Check for proper connection
Worn or loose footswitch socket
Return to supplier
Faulty stimulator
Return to supplier
If there is no stimulation during gait, remove
footswitch
from the shoe and try to trigger
between 2 fingers. If this triggers the stimulator, consider moving the position of the
footswitch
in the shoe to a more suitable location. Refer to page 34.
2. No stimulation but the LED light flickers and the screen indicates stimulation
Fault
Action
Faulty electrode lead
Replace electrode lead
Faulty electrodes
Replace electrodes
Lead not fully inserted/connected
Check for proper connection
Worn or loose electrode socket
Return to supplier
Faulty stimulator
Return to supplier
Impact or water damage
Return to supplier
It is easier to test whether the stimulator is delivering an output by pressing the
test button
rather than relying on the
footswitch
.
3. Incorrect movement produced by the stimulation
Fault
Action
Incorrect electrode placement
Refer to electrode placement instructions
Poor electrode contact
Dampen the electrodes and skin with water
Muscle fatigue
Rest
It could be that the user’s medical condition has changed recently and therefore alternative
settings may be required.