Hunan Next Generation Instrumental T&C Tech. Co., Ltd.
NGI
l
NGI
N39200 Series
High-accuracy Dual-channel Programmable DC Power Supply
48
No output.
Table 22
Fault
Possible Causes
Solutions
There is no output after
switching on.
The
voltage/current
setting is zero.
Set
the
desired
voltage/current.
The voltage output stops
after transient output.
There is over voltage
protection (OVP).
Lower the output setting.
It is not allowed to set output voltage/current properly.
Table 23
Fault
Possible Causes
Solutions
The
output
voltage/current cannot be
set properly.
The set voltage/current is
higher
than
V-Limit
Max/I-Limit Max.
Set
the
desired
voltage/current.
The set voltage/current is
lower
than
V-Limit
Min/I-Limit Min.
Unstable output voltage.
Table 24
Fault
Possible Causes
Solutions
The output voltage are not
stable.
N39200
is
switching
between CV and CC.
Change the setting.
The output voltage is
wobbly.
The sense terminal is not
connected.
Connect
to
sense
terminal, by using the
4-pin green connector
supplied.
High output voltage ripple.
Table 25
Fault
Possible Causes
Solutions
The
ripple
is
high
sometimes.
The output voltage is out
of range.
Adjust the output voltage.
The ripple is high in
another
operation
environment.
There
is
strong
electromagnetic
interference nearby.
Stay away from the
interference.