7.3 Performance tests
WF1943B
7-7
Amplitude vs. frequency characteristics
Connection:
Use coaxial cable to connect FUNCTION OUT to AC voltmeter 2 (50
Ω
termination).
Setting:
Initialize, then set the amplitude to 20 Vp-p/open and select the frequency and
waveform as indicated in the table.
Measurement:
Measure the true rms output voltage for each frequency and waveform.
Judgment:
The normal ranges are indicated in the table.
Waveform
1 kHz
Setting
to 500 kHz
to 1 MHz
to 3 MHz
to 10 MHz
to 15 MHz
(Reference value)
+0.2/-0.3 dB
+0.2/-0.3 dB
+0.35/-0.7 dB
+0.5/
−
1.5 dB
+0.5/
−
2.0 dB
(Reference value)
±
0.3 dB
――――
――――
――――
――――
(Reference value)
±
0.3 dB
±
0.3 dB
――――
――――
――――
(Reference value)
±
0.5 dB
――――
――――
――――
――――
(Reference value)
±
0.5 dB
――――
――――
――――
――――
Sinewave distortion
Connection:
Use coaxial cable to connect FUNCTION OUT to a distortion meter (50
Ω
termination).
Setting:
Initialize, then set the amplitude to 20 Vp-p/open and the frequency as indicated in the
table.
Measurement:
Measure the distortion
Judgment:
The normal range is indicated in the table.
Frequency Normal
range
10 Hz to 100 kHz
0.2 % or less
(Bandwidth 500 kHz)
Содержание WF1943B
Страница 2: ......
Страница 3: ...DA00012503 002 WF1943B MULTIFUNCTION SYNTHESIZER Instruction Manual...
Страница 4: ......
Страница 8: ......
Страница 83: ...4 2 Sweep WF1943B 4 35 Sweep value and SYNC OUT Approximately 100 to 125 s...
Страница 150: ......
Страница 152: ......
Страница 154: ......
Страница 155: ......
Страница 156: ...http www nfcorp co jp NF Corporation 6 3 20 Tsunashima higashi Kohoku ku Yokohama 223 8508 JAPAN...