AUTOCLAVE:
1) Brush off excess dirt and wipe with an alcohol-soaked cloth.
2) Remove the handpiece from the dental unit tube and lubricate with spray oil or other standard oil as described under 5. Lubrication.
3) Insert the handpiece into a sterilization bag and seal it.
4) The sterilization of the drill key is the same as that of the handpiece.
5) Sterilize up to a maximum of 134°C, Symbol Sterilization .Do not set the sterilization temperature higher than 134°C.
6) Please dry the handpiece by 134 ºC dryer (15 minutes) after high temperature sterilization and disinfection.
7) The product should be high temperature sterilized and used on site.
8) The handpiece is resistant to at least 250 times of high temperature sterilization and disinfection.
PREPARATION:
1) Brush off excess dirt and wipe with a cloth soaked in alcohol.
2) Remove the handpiece from the dental unit tube and lubricate spray lubricant or other spray oil as described in the Lubrication section.
3) Place the handpiece in a sterilization pouch and seal it.
4) Sterilize up to a maximum of 134 °C.
CAUTION
The heating element may be located at the bottom of the chamber, and the temperature may exceed the set value. To prevent this, place
the handpiece in the center or top tray.
Содержание Rose 202-M4 2SH
Страница 68: ...68 ...
Страница 73: ...73 2 PARÁMETROS TÉCNICOS 1 Motor de aire de agua interno 2 Pieza recta ...
Страница 86: ...86 2 Straight handpiece 3 Contraangle handpiece ...
Страница 98: ...98 2 PARAMÈTRES TECHNIQUES 1 Moteur interne eau air 2 Pièce droite ...
Страница 111: ...111 2 PARÂMETROS TÉCNICOS 1 Motor de ar aquático interno 2 Peça recta ...
Страница 124: ...124 2 PARAMETRI TECNICI 1 Motore ad aria con acqua interna 2 Pezzo diritto ...