9
EDH0185En1020 — 07/11
VP-25XA
Precision Compact Linear Stage
6.2
Connection
On the stage is a label which indicates its name, its serial number and the
motor it is equipped with (eg.: UE25CC).
WARNING
Always turn power OFF on electronics units before connecting them to
stage.
Stages may be connected to the rear panel motor connectors labeled
“Motor…” any time prior to power-up with the supplied cable assemblies.
WARNING
With MM series controllers, damage to stage may occur if the stage is not
the same type as shown on driver label located near the stage interface
connector.
Check that the option number specified on this label correspond to the
number indicated in the driver module options table for your stage.
6.3
Cables
Our VP-25XA stage is delivered equipped with a 1.5-meter cable with 25-pin Sub-
D connector. It can be directly connected to our controllers/drivers of MM or
ESP series.
WARNING
These cables are shielded correctly. For a correct operation, make sure
to lock connectors (ground continuity provided by cables).
WARNING
Keep the motor cables at a safe distance from other electrical cables in
your environment to avoid potential cross talk.
VP-25XA
S/N
#
Encoder:
5V
Motor:UE
25
CC
D.C
Motor
U=48
V
DC
I=1
A
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com