90
Remote Command Set
ZZ
Description
Configuration register set
Syntax
ZZ
nn
Remarks
This command is used to configure the default behavior of some of the controller’s
features. It is typically followed with an SM (Save to Memory) command.
Bit#
Value
Description
*0
0
Perform auto motor detection. Check and set motor type
automatically when commanded to move.
0
1
Do not perform auto motor detection on move
*1
0
Do not scan for motors connected to controllers upon reboot
(Performs ‘MC’ command upon power-up, reset or reboot)
1
1
Scan for motors connected to controller upon power-up or reset
Example:
ZZ 1; SM
(Disable automatic motor check and save to memory)
ZZ 3; SM
(Disable motor check on move, check motors on system reset)
See Also
*RCL, MC
ZZ?
Description
Configuration register query.
Syntax
ZZ?
Remarks
This command is used to query the configuration register setting.
Response
Value
Description
nn
Integer
Configuration register.
Example:
ZZ ? (Get configuration register settings)
1
(Returns 1; bit-0=1 & bit-1=0; means auto motor detection on move is disabled)
See Also
ZZ
Содержание New Focus Picomotor 8742
Страница 1: ...Picomotor Controller Driver User s Manual Model 8742 ...
Страница 53: ...Computer Interfacing 53 Cycle Tab Cycle Motion ...
Страница 54: ...54 Computer Interfacing Terminal Tab Command Line Terminal ...
Страница 57: ...Computer Interfacing 57 Options Window Connection Options Setup ...
Страница 58: ...58 Computer Interfacing Properties Window USB and Ethernet Properties ...
Страница 59: ...Computer Interfacing 59 Controller Scan Scan RS 485 LAN ...
Страница 60: ...60 Computer Interfacing Address Conflict Resolution Window Automatic Address Conflict Resolution ...
Страница 61: ...Computer Interfacing 61 Device Address Window View Change Controller Address ...