April 2018
54
FPM-8220
MODE:DBM
Common
Device Dependent
Front Panel
Action
Selects Logarithmic (dBm) units to be used in reporting measurements.
Examples
Mode:dBm
- Measurements will be reported in dBm.
MODE:W
Common
Device Dependent
Front Panel
Action
Selects Linear (watts) units to be used in reporting measurements.
Examples
Mode:W
- Measurements will be reported in Watts.
MODE?
Common
Device Dependent
Front Panel
Action
Requests FPM-8220's measurement mode.
Response
One of the following ASCII character strings:
dB
dBm
W
Notes
This information is available on front panel indicators. See Chapter 2.
The response to MODE? is the units of value that will be used for responses to a POW? request.
Example
s
Mode? - Response W means that linear measurement mode (watts) is in effect.
POWer?
Common
Device Dependent
Front Panel
Action
Requests the most recent value of measured optical power.
Response
Numeric value.
Notes
Units defined by MODE commands. Value is returned when the next measurement is available in
slow mode. It may take up to 5 seconds.
In dB or dBm mode, reports the same number as on the display.
In W mode, reports in scientific notation.
Monitor the Device Event Status Register for over-range or under-range conditions when using this
query.
Examples
Mode?;Power? - Response of DBM,-13.584 means the last measured power was-13.584 dBm.
MODE:W;POW? - Response 2.795E-006 means the last measured power was 2.795
μ
W.
Содержание FPM-8220
Страница 1: ...User s Guide Fiber Optic Power Meter FPM 8220 70044001 April 2018...
Страница 2: ......
Страница 6: ...April 2018 iv FPM 8220 This page was intentially left blank...
Страница 12: ...April 2018 x FPM 8220 This page was intentionally left blank...
Страница 16: ...April 2018 4 FPM 8220 Specifications...
Страница 34: ...April 2018 22 FPM 8220...
Страница 78: ...April 2018 66 FPM 8220...