Math Functions
1919-R User Manual
75
9.7
Density
You can display power/energy as measured by the sent as W/ cm² based on user-defined laser
spot shape and size.
Note:
The
Density
option is grayed out in any of the following conditions: if the present range is
dBm, if
Normalize
has been set, or if Exposure is the current mode.
To display as power/energy density:
1.
Press
Menu
and enter
Functions
.
2.
Go to
Density
. Press the
Enter
key.
3.
Select the beam shape (
Disabled
,
Rectangle
,
Circle
). Selecting
Rectangle
enables the
Height
and
Width
parameters for modifying the spot height. Selecting
Circle
enables the
Diameter
parameter for modifying the spot diameter.
Figure
9-8 Modifying Density – Circular Beam and Spot Diameter
4.
After the shape (
Rectangle
or
Circle
), select the size (
Height
and
Width, or Diameter
). Size
parameters can range from 0.1mm to 100.0mm.
The measurements will be shown as cm² on the graphical display on the main measurement
screen.
Figure
9-9 Measurement of Power Density in nW/cm
2