1919-R User Manual
107
13
1919-R Specifications
Topics include:
13.1
System/Meter Specifications
Table
13-1 System/Meter Specifications
Input Specifications
Thermopile, Photodiode
Input Ranges
15nA - 1.5mA full scale in 16 ranges
A to D Sampling rate
15Hz
A to D resolution
18 bits plus sign
Electrical accuracy
±0.25% ± 20pA new; ±0.5% ±50pA after 1 year
Electrical input noise level
500nV or 1.5pA + 0.0015% of input range @3Hz.
Dynamic range
9 decades (1:109)
Input Specifications
Pyroelectric Sensors
Input Range
0 - 6V full scale
A to D Sampling rate
5000 Hz
A to D resolution
12 bits no sign (0.025% resolution)
Electrical accuracy
±0.25% new; ±0.5% after 1 year
Electrical input noise
2mV
General Specifications
Detector Compatibility
Thermopile (919P series), Photodiode (918D, 818DB, 819C/D series),
and Pyroelectric (919E series)
PC Interface
USB; RS232 (max baud rate 115200)
Analog output
1v full-scale; 0.03% resolution. 100 ohms impedance
Analog output accuracy
±0.2% (of reading) ±0.3% of full scale volts
Dimensions
114W x 41D x 212H
Mass
470g
Display
320x240 pixel TFT LCD; Active area 70x52mm approx.
Display digit height
15mm
LCD lighting
LED’s. Operates from charger or battery. Lighting level can be
adjusted between 3 levels using on/off button.