Graphical Displays
1919-R User Manual
64
8.7.1
Modes
The only
Mode
that Exposure displays can be used with is:
Exposure
8.7.2
Manual/Timeout/Pulses
Press
Manual
(the rightmost soft key) to display the Exposure End menu. Use the up
↑
/down
↓
and left
←
/right
→
Navigation keys to select the options to end the exposure measurement:
Option
Description
Sample Figure
Manual
Stops exposure measurement by the
user pressing the
Stop
key.
Figure
8-27 Manual Option Selected
Stops after the specified time has
elapsed (as shown on the key legend).
You can select: 1 second, 3 seconds, 10
seconds, 30 seconds, 1 minute, 3
minutes, 10 minutes, 30 minutes, or 1
hour. Will also stop by the user pressing
the
Stop
key.
Figure
8-28 Timeout Options
Stops after the specified number of
pulses has arrived (as shown on the key
legend). You can select: 1, 2, 5, 10, 20,
50, 100, 200, 500, 1000, 2000, or 5000.
Will also stop by the user pressing the
Stop
key.
Figure
8-29 Pulse Options
8.7.3
Exposure for Photodiode Sensors
Exposure can be measured with Photodiode sensors as well. The following are the differences
between measuring Exposure with Pyroelectric Energy Sensors and Photodiode Sensors:
Photodiode sensors actually measure power and sum the results into energy.
There is no After Pulses option with Photodiode Sensors.