background image

11

3.3.1  Detector Calibration and Accuracy

Newport Corporation calibrates its detectors using secondary standards
directly traceable to the United States National Institiute of Science and
Technology (NIST) or to Great Britain’s National Physical Laboratory, (NPL).
The details and accuracy of the calibration procedure vary with each detector
model but a detailed description of the calibration results is supplied with
each individually calibrated detector.

In general, detector calibration accuracy ranges from 2% to 5% in absolute
terms and varies with wavelength.  Each detector will have some variation in
the response over different sections of its surface. Therefore, for the most
reproducible measurements, light should illuminate the detector as uniformly
as possible over 80% of the detector’s active area.

Caution

Avoid focusing a light source onto the detector surface.  Inaccurate
readings and possible detector damage may result.  Consult your
detector manual for information on detector saturation or damage
thresholds.

NIST traceability requires that detectors be recalibrated on one year intervals.
As individual detector responses change with time, especially in the ultravio-
let, recalibration is necessary to assure confidence in the accuracy of the
measurement.  For the most reproducible measurements, the same detector
should always be used for measurements which are to be directly compared.

3.3.2  Quantum Detector Temperature Effects

Semiconductor based photodiode detector characteristics (Newport Corporation

Low-Power

 detectors) are significantly affected by temperature.  At longer

wavelengths, these quantum detectors typically lose sensitivity with increasing
temperature.  However the detector dark current increases exponentially with
temperature.  For silicon detectors, dark current is generally on the order of a few
picoamps at room temperatures.  With uncooled germanium detectors, however,
this dark current is on the order of a nanoamp, or typically 1,000 to 10,000 times
greater than silicon.  These dark currents can be zeroed at any moment in time via
the Model 1815-C’s ZERO knob.  Since dark currents drift with temperature, the
ZERO should be adjusted just prior to taking any measurements.  The noise or
drift in the dark current sets a lower bound on the measurement resolution which
can be achieved with any given detector.

If the detector temperature is constant, sensitivity changes and dark current
drifts are significantly reduced.  In addition, if the detector is cooled, the dark
current and dark current noise will decrease.  For the most accurate measure-
ments, particularly with germanium detectors, the user should cool the
detector to approximately 0 

°

C and control the temperature to within 

±

 1 

°

C.

3.3.3  Thermopile Detector Temperature Effects

Disk themopile based detector characteristics (Newport Corporation 

High-

Power

 detectors) are significantly affected by temperature fluctuations arising

from air flow disturbances.  As the detector element is a heat measuring
device, air flow disturbances often set a practical lower limit on the power that
a detector can measure.  In order to get the most out of any thermopile
detector, be careful to shield the detector from air flow disturbances.  Com-
mon sources of disturbance are air conditioners and people walking past.

Содержание 1815-C

Страница 1: ...Model 1815 C Optical Power Meter OPERATOR MANUAL...

Страница 2: ......

Страница 3: ...anty This warranty does not apply to defects resulting from modification or misuse of any product or part This warranty also does not apply to fuses batteries or damage from battery leakage This warra...

Страница 4: ...r EN55011 Group 1 Class A EN50082 1 IMMUNITY Electrostatic Discharge per IEC 1000 4 2 severity level 3 Radiated Emission Immunity per IEC 1000 4 3 severity level 2 Fast Burst Transients per IEC 1000 4...

Страница 5: ...Using the System 5 2 4 1 Battery Installation Replacement 6 2 4 2 Power Up 6 2 4 3 Detector Connection and Setup 6 2 4 4 Setting the Detector Calibration Factor 7 2 4 4 1 Setting the Calibration Facto...

Страница 6: ...2 Performance Verification 14 4 2 1 Environmental Conditions 14 4 2 2 Recommended Test Equipment 14 4 2 3 Initial Conditions 15 4 2 4 Verification Procedures 15 4 3 Troubleshooting Guide 17 Section 5...

Страница 7: ...iagram 10 Figure 7 Model 1815 C Range Switch Positions 15 List of Tables Table 1 Model 1815 C Compatible Detectors 2 Table 2 Model 1815 C Controls 3 Table 3 SETUP DIP Switch Functions 5 Table 4 SETUP...

Страница 8: ...able to this instrument Warning The American National Safety Institute ANSI states that a shock hazard exists when probes or sensors are exposed to voltage levels greaterthen42VDCor42VpeakAC Donotexce...

Страница 9: ...pe C degrees Centigrade DC direct current F degrees Fahrenheit Hz hertz cycles per second I V current to voltage converter kHz kilohertz k kiloOhm mA milliamp mV millivolt nA nanoamp nF nanofarad nm n...

Страница 10: ...5 Digit LCD Sample Rate 2 5 Hz Range Switch 6 positions one per decade of gain Operating Environment 70 RH Noncondensing 18 C 28 C 65 F 82 F Storage Environment 90 RH Noncondensing Compatible Detector...

Страница 11: ...the calibration factor in the following way Display Noise Floor specified noise floor calibration factor mantissa Voltage Amplification High Power Detectors Range Switch Position 3 4 5 6 High Power Fu...

Страница 12: ...floor is greater than specified see Section 3 3 2 Quantum Detector Tempera ture Effects 6 ACCEL ADJ fully clockwise worst case Detector Responsivity Limits Detector in Use Allowed Responsivities Allo...

Страница 13: ...xii...

Страница 14: ...FP3 FH1 818T 10 818T 30 818T 150 818T 150X 818T 300 High Power Low Power W mW W W W W W mW mW mW W W W kW 20 200 2 2 200 20 nW nW LOW POWER HIGH POWER LOW POWER W ATTN CAL ADJ CAL POWER ZERO ON 1 OFF...

Страница 15: ...del 1815 C Optical Power Meter Refer to Figure 1 for where each accessoryfitswithintheModel1815 COpticalPowerMeterfamilytree For moreinformation pleaserefertotheNewportCatalog Table1 Model1815 CCompat...

Страница 16: ...The front and rear panels of the Model 1815 C are shown in Figures 2 and 3 respectively ThefollowingtablegivesashortdescriptionofeachModel 1815 Ccontrol Table2 Model1815 CControls Control Description...

Страница 17: ...Adj Reqd See Manual Cal Factor Exp Low Power Low Power w Attn High Power INPUT 1 0 ANALOG OUTPUT ACCEL ADJ DIP Switch Input BNC Connector Acceleration Time Constant Adjust Output BNC Connector SETUP...

Страница 18: ...h Thefollowingtableidentifiesthefunc tioneachSETUPswitchpositionperforms Table3 SETUPDIPSwitchFunctions SETUPSwitch SwitchPosition Description 1 2 SelectsforLow Power semiconduc tor detector 1 0 orHig...

Страница 19: ...themeter Thedigitaldisplayshouldlightindicatingthatthemeteris operating IftheBATannounciatorislit replacethebatteries SeeSection 2 4 1 2 4 3 Detector Connection and Setup Connect the detector to the I...

Страница 20: ...ed ddofthefullcalibrationfactor valued ddE d Themantissaisthemostfrequentcalibrationadjustment madewhenmakingmeasurementsatvariouswavelengths Tosettheman tissa usethefollowingprocedure 1 Pressthefront...

Страница 21: ...econdtimeconstant theACCELADJ potentiometerisadjustedatthefactoryforthisvalue However becausethe actualtimeconstantisnotexactly1secondforallthesmallerthermopiles it mayberequiredtooptimizetheaccelerat...

Страница 22: ...no adjustment to the ACCELADJisrequired Note AccelerationcanonlybeusedwithHigh Power thermopile detec tors Ifaccelerationisusedwithotherdetectors erroneousreadings canresult 2 4 6 Analog Output The Mo...

Страница 23: ...ordiskthermopiles voltagemeasurement DetectorsignalsareintroducedtotheModel1815 CbywayofaBNCinput connector A bank of DIP switches found on the back of the Model 1815 C configurestheamplifieraseithera...

Страница 24: ...ion Low Powerdetectors aresignificantlyaffectedbytemperature Atlonger wavelengths thesequantumdetectorstypicallylosesensitivitywithincreasing temperature Howeverthedetectordarkcurrentincreasesexponent...

Страница 25: ...Powerdiskthermopiledetectors whichisnotcomingfromthesource andturningoffroomandotherlights Note Changes in ambient light levels can occur from such factors as turning room lights on or off or by movin...

Страница 26: ...easurementwas made WavelengthCalibration Any CheckthattheCalibration Factorforthemeasure mentwavelengthis properlyset LowBatteryPower Any Checkthatmeterisnot indicatinglowBAT DetectorSaturation Low Po...

Страница 27: ...ion Note If the instrument does not meet specifications and it is still under warranty lessthan12monthssincedateofshipment contactyour local Newport Corporation representative or the factory to deter...

Страница 28: ...gapreciselysetablecurrentsourcesuchasthe KeithleyInstrumentsModel220andapreciselysetableVoltageSourcesuchas theHewlett Packard8904A FunctionGenerator W W kW W W mW W nW nW mW W W W 2 200 20 20 200 2 m...

Страница 29: ...htoposition3 thehighestgainpositioninthe voltage mode See Figure 7 3 PresstheCALbuttonandadjustCALADJuntilthedisplayindicates 1 00andthenreleasetheCALbutton 4 With a shorted input use the ZERO knob to...

Страница 30: ...hissection pleaseseeSection5fordetails onreturningyourentiresystemtoNewportforservice Table10 Symptom FaultTroubleshootingGuide Symptom PossibleFault Correction Blankdisplay PowerswitchOFF TurnswitchO...

Страница 31: ...anbatteries see Section2 4 1 Itscalibrationaccuracyiswarrantedforaperiodof1year After 1year theunitshouldbereturnedtoNewportCorporationforrecalibration Toobtaininformationconcerningfactoryservice cont...

Страница 32: ..._________________________________________ Company ______________________________________________________________________________ Address _______________________________________________________________...

Страница 33: ...20...

Страница 34: ...en certified compliant with ISO 9001 by the British Standards Institution P N 19571 01 Rev F IN 03923 1 00 Printed in the USA Newport Corporation Worldwide Headquarters 1791 Deere Avenue Irvine CA 926...

Отзывы: